Project

General

Profile

Actions

Feature #10612

closed

Use precalibrated ctf estimate for tomography defocus setting

Added by Anchi Cheng almost 3 years ago. Updated over 1 year ago.

Status:
Closed
Priority:
Normal
Assignee:
Category:
-
Target version:
Start date:
07/26/2021
Due date:
% Done:

0%

Estimated time:
Deliverable:

Description

Today's goniometer should be reproducible enough.

Actions #1

Updated by Anchi Cheng over 2 years ago

1. Collect tilt series using data collection settings on specimen that gives strong Thon rings. Exposure time can be adjusted to give good image.
2. If this is a movie-capable camera, align the frames to get better images.
3. estimate CTF using appion
4. Download CTF data of the the tilt series using myamiweb
5. Run leginon/tomography/upload_ctf.py to upload the ctf value difference as calibration.
6. If there are calibration at the given tem instrument, it will be used as extra correction on top of the UCSF model.

Actions #2

Updated by Anchi Cheng over 1 year ago

  • Status changed from Assigned to Closed

Tested with success, but not widely used.

Actions

Also available in: Atom PDF