Project

General

Profile

Actions

Bug #8467

closed

K3 needs to use prepareDarkReference to make dark current correction, and need it more often, too

Added by Anchi Cheng over 4 years ago. Updated over 4 years ago.

Status:
Closed
Priority:
Normal
Assignee:
Category:
-
Target version:
Start date:
01/10/2020
Due date:
% Done:

0%

Estimated time:
Affected Version:
Appion/Leginon 3.4
Show in known bugs:
No
Workaround:

Description

Noticed K3 image correction goes bad quicker than K2 but AcquireHardwareDark command is not available for K3. Learned that it is replaced by a different function. It probably also need to be done more often than N2 fill time since it runs faster.

Actions #1

Updated by Anchi Cheng over 4 years ago

  • Status changed from New to Assigned

f0f3e4bb should be linked to this issue, too.

Actions #2

Updated by Anchi Cheng over 4 years ago

added option to acquire dark current reference more often.

The condition checks at each repeat time. If extra dark current reference is required, and the last time it is acquired is longer than "dark current reference repeat time", then it will acquire dark reference even though autofiller does not require refill.

Actions #3

Updated by Anchi Cheng over 4 years ago

  • Status changed from Assigned to Closed
  • Assignee set to Anchi Cheng

Tested. The number of dark references acquired at each refill is close to enough. Some users may prefer increasing the frequency.

Actions

Also available in: Atom PDF