Project

General

Profile

19 Å mouse mFth1 recombinant apoferritin image-shift data collection with image-shift aberration-correction » History » Version 5

Anchi Cheng, 03/21/2019 06:41 PM

1 2 Anchi Cheng
h1. 1.9 Å mouse mFth1 recombinant apoferritin image-shift data collection with image-shift aberration-correction 
2 1 Anchi Cheng
3
This data collection correct the coma and stigmatism that is linearly dependent on the distance of beam-image shift.  It allows quality approaching stage shift approach but even higher image-shift image range than previously acceptable.  In this particular setup and microscope, up to 36 1.2/1.3 gold foil holes can be taken in one hl image that requires one auto focus. The pixel size used is significantly smaller than previous setup shown to increase S/N ratio at the same resolution.  FSC 0.143 is at 1.92 A.
4
5
The resolution and speed gain relative to previous image-shift data collection is from
6
7
# single construct of the recombinant apoferritin and better purification.
8
# beam-image-shift aberration correction allows larger shift at the same quality of data collection as stage shift but a fraction of time in stepping through.
9
10
*IMPORTANT*: The number of holes can be imaged depend on the scope-camera configuration. Some scope may show too much image distortion at low magnification for accurate targeting.
11
12
h2. Instruments
13
14
SEMC@NYSBC Krios1
15
# Scope: FEI Titan-Krios, 300 kV, micro probe at low mag, nano probe at high mags
16
# Camera: Gatan K2. mount below Falcon/Ceta. 0.66 Å at 37kx scope nominal mag
17
18
h2. Grid
19
20
* 1.2/1.3 C-flat carbon support coated with gold; 400 mesh Copper grid.
21
22 5 Anchi Cheng
{{include(Krios1-K2-NYSBC-large_shift)}}
23 1 Anchi Cheng
24
h2. Application:
25
26
MSI-T2 version 3.4 and up (This is similar to older version of MSI-T2 advanced).
27
28
h2. Presets Manager:
29
30
No preset cycling (Krios has its own normalization).
31
32
h2. Queuing usage:
33
34
Hole Targeting: Queuing, target at the carbon area to cover four holes.
35
Exposure Targeting: *No Queuing*. automated targeting.
36
37
h2. Move method to reach the targets:
38
39
Hole: Presets Manager stage shift
40
Focus:  Presets Manager image shift
41
Exposure: Presets Manager image shift.  Turn on "Correct image shift coma effect" in the Advanced Settings.
42
43
h2. Wait time before final exposure chosen to reduce maximal frame drift to average to 2 Angstroms :
44
3 seconds
45
10 seconds addition for the first target
46
47 3 Anchi Cheng
h2. N2 Filling node settings
48 1 Anchi Cheng
49 3 Anchi Cheng
Need to check more often and increase the threshold so that it does not start filling during the 10 minute (25-30 movies) long exposure targeting per hl.
50
!large_image_shift_autofiller_settings.png!
51
52
h2. [[caling Stage Position Matrix Calibration|A trick to get Image shift targeting accuracy better]]