Beam Shift matrix calibration » History » Version 10

Anchi Cheng, 05/15/2012 07:26 PM

1 10 Anchi Cheng
h1. Beam Shift matrix calibration (Optional)
2 1 Amber Herold
3
4
5
Beam shift matrix calibrations are used at each magnification that Leginon uses. These
6
calibrations are not absolutely necessary, but can be very helpful to center the beam in the
7
navigator node. In this case, the image of a small beam within the CCD is treated as an object
8
in the image. When the beam is shifted by the electromagnetic lenses of the microscope, the
9
imprint of the beam moves relative to the CCD. The 2x2 transformation matrix created then
10
relates the values sent to the microscope and the amount of "beam shift" movement seen on the
11
CCD imaging area. Only one set of measurement is sufficient for beam shift matrix
12
calibration.
13
14
15
16
17 8 Amber Herold
[[Image Shift matrix calibration|How does matrix calibration work?]]
18 1 Amber Herold
19
20 3 Alex Kadokura
21 9 Amber Herold
# Leginon/Presets Manager> Select a preset for the calibration and send its
22 1 Amber Herold
parameter to the microscope. Matrix calibration depends only on magnification and
23
microscope high tension. Therefore, only one preset per combination needs to be
24
calibrated.
25 9 Amber Herold
 
26
#  Scope> contract the beam sufficiently so that it remains in the area of CCD
27 1 Amber Herold
acquisition during the calibration. It is also preferable that the CCD is imaging an
28
area with no distinct feature such as an empty grid square to minimize false peak from
29
unmoved object during a beam shift. At low magnifications, stage can be pulled out and
30
parked with a spacer to create the empty field.
31 9 Amber Herold
 
32
#  Leginon/NodeSelector> select "Matrix" node.
33
 
34
#  Leginon/Matrix> the preset CCD configuration is almost certainly bad for imaging
35 1 Amber Herold
contracted beam. Therefore, DO NOT use "test acquire" at this point.
36 9 Amber Herold
 
37
#  Leginon/Matrix/Toolbar> open "settings" window by clicking the icon to select
38 1 Amber Herold
camera configuration and correlation method. The former will take into effect only if
39
"Overwrite Preset" is checked. Click "OK" to save the settings and close the window when
40
done.
41 9 Amber Herold
 
42 1 Amber Herold
*Tip: 1024x1024 binned by 4 and a VERY short exposure time can save time but still
43
keep the contracted beam from burning the CCD.
44 9 Amber Herold
 
45
#  Leginon/Matrix/Toolbar> left-click "Acquire Image" to obtain a test image with
46 1 Amber Herold
current parameters.
47 9 Amber Herold
 
48
#  Leginon/Matrix/Toolbar> select "beam shift" as the Parameter and open the
49 1 Amber Herold
"Parameter Setting" window by clicking on the icon to the right of the selector.
50 9 Amber Herold
 
51
#  Leginon/Matrix/Matrix Settings> "Average # position"=1 is sufficient. The rest
52 1 Amber Herold
can be left in default values. "Interval" is not a relavent parameter since " average #
53
position"=1.
54 9 Amber Herold
 
55
#  Leginon/Matrix/Toolbar> left-click (Execute icon) to calibrate.
56
 
57
#  The image of the beam should be shifting 10-30% of the imaging area. If this is not
58 1 Amber Herold
the case, then adjust shift fraction so that this occurs. The images can be monitored in
59
Image Display Panel with display selection in image control panel set to "image". The
60
cross correlation and its peak can also be displayed. The beam need to be contained in
61
the imaging area at all time.
62 9 Amber Herold
 
63
#  Use Navigation node to [[Checking Matrix and Modeled Stage Position Calibration|check the result of the calibration]].
64 1 Amber Herold
65
66
67
68 9 Amber Herold
*Beam Shift Calibration Need for the Example MSI:*
69 1 Amber Herold
70
71
|<filename>Preset</filename>|<filename>magnification</filename>|
72
|gr|120 (rarely used)|
73
|sq|550 (rarely used)|
74
|hl|5000 (seldom used)|
75 7 Anchi Cheng
||11000 (useful in beam shift alignment tool in Presets Manager to align the beam at
76 1 Amber Herold
50000x)|
77
|fc,fa,en,ef|50000 (useful)|
78
79
80
81
82 8 Amber Herold
*The Presets Manager beam shift alignment tool works best using ~5x lower magnification than the
83 1 Amber Herold
preset meant to be aligned in order to see the whole beam in view. This means that for
84
fc,fa,en,ef that are at 50000x, a calibration at 11000x is required at the similar defocus
85
range.
86 2 Amber Herold
87
______
88
89
[[Image Shift matrix calibration|< Image Shift matrix calibration]] | [[Stage Position matrix calibration|Stage Position matrix calibration >]]