Project

General

Profile

Actions

Check Phase Plate Patches » History » Revision 4

« Previous | Revision 4/6 (diff) | Next »
Anchi Cheng, 02/15/2017 03:06 PM


Check Phase Plate Patches

imaging the phase plate slot at low magnification to check for contamination and recovery of the irradiated spots

On FEI Volta phase plate slot there are 76 patches to be used. Each can be accessed by "Next" button in TEM user interface (TUI). Once Leginon is set up with AutoIt to click this button, it is possible to go through all patches and record diagnosis images which is the purpose of this node.

Grid used: Grating Replica

  1. Remove Phase Plate, align the scope to eucentric focus, make sure the grid is at the eucentric height, and adjust imaging condition to give ~ 1 Å pixel. Exposure time should be enough to give clear Thon ring pass 4 Å resolution after direct detector frame alignment. Save this condition as the preset to be enterred in "Check PP Patches" node settings. The defocus should be -0.5 um and free of objective astigmatism. It should also use full resolution and size of the camera.
  1. Insert Phase Plate, wait for it to settle.
  1. Adjust to on-plane condition for the phase plate with the same preset as above.
  1. In "Calibration Application", select "Check PP Patches" node and open settings to add the preset to its preset order. Also increase "Wait time before acquiring image" to at least 5 second.
  1. Confirm on TUI under Aperture tag that the "Next" button is visible for AutoIt to click.
  1. Click on "Simulate Target" tool in "Check PP Patches" node. A dialogue should appear to allow you to enter phase plate identity, the current patch position etc. Once "OK" is clicked in the dialog, the acquisition will start.

The result images and its phase plate identity is saved. You can view them in the myamiweb image viewers.

Generate Phase Plate Condition Map

Updated by Anchi Cheng over 7 years ago · 4 revisions