h1. Good Alignments Save Time
2
3
Leginon relies on a stable microscope. The most likely reason for leginon to fail performing is bad microscopy alignment. Various bad alignments often compound the difficulty in problem solving. The requirement of leginon for good alignment is higher than most users who set up low dose image acquisition directly at the scope are used to, especially regarding alignment in LM mode since most low dose user has limited usage for LM mode images.
4
5
6
Apart from alignments that every user does without Leginon, Here are some of alignment critical to Leginon MSI applications but not a common low-dose kit users.
7
8
9
10
h2. HM beam-tilt pivot point and current rotation center set for eucentric focus
11
12
13
14
These are needed for beam-tilt based auto-focusing
15
16
h2. Defocus reset to eucentric at BOTH HM and LM modes
17
18
These are important starting point to allow consistent performance of calibrations and saved presets.
19 7 Anchi Cheng
20 6 Anchi Cheng
If your scope control panel has a "go to eucentric focus" button, and the set point is no more than 3 um off in HM and 10-20 um off in LM, it would be good enough to just press that and then reset defocus.
21 1 Amber Herold
22
What is [[eucentric focus]]?
23
24
h2. Image/Beam Calibration at HM mode
25
26
27
28 4 Anchi Cheng