Project

General

Profile

RCT run protocol from a user » History » Version 4

Anchi Cheng, 10/17/2012 06:02 PM

1 1 Anchi Cheng
h1. RCT run protocol from a user
2
3
h2. Setup
4
5
h3. Days before
6
7
* Pump down stage
8
*   Cryo-cycle for 4 hours (240 min)
9
*   Prepare grids 
10
11
h3. Scope alignments
12
13
2.2.1 no sample
14
15
    gun shift/tilt
16
    condenser aperture
17
    condenser stig.
18
    check if cold trap is in
19
    adjust eye pieces 
20
21
2.2.2 with sample
22
23
    z-height
24
    focus both M and SA modes
25
    objective stig.
26
    objective aperture at hi mag, use focus to shrink beam, not intensity
27
    at zero defocus
28
        beam tilt pivot points X,Y
29
        rotation center 
30
31
h3. Presets and Magnifications
32
33
|preset|mag|camera config|defocus|notes|
34
|gr|120X|512x8|0|do not cycle through|
35
|sq|1700X|1024x4|-3.0e-4|for square_centering|
36
|cs|1700X|1024x4|-3.0e-4|for rct_targeting, copy of sq|
37
|hl|5000X|512x8|-5.0e-5|image shift alignment and z-focus|
38
|en|50kX-100kX|4096x1|-1.5e-6 to -3.5e-6| |
39
|fa|50kX-100kX|512x4|-1.5e-6|beam focus and drift checks|
40
41
h3. Leginon
42
43
2.4.1 pre-Altas
44
45
    beam shift/intensity for each preset
46
    align image shifts btw presets
47
    matrix calibrate, stage position for gr preset
48
    optional: fine tune rotation center in Leginon, available in any focus node using the fa preset
49
    on rare occasions: check optical axis using Leginon tool, this is only required if the square is shifting alot when you tilt the stage 
50
51
2.4.2 post-Atlas
52
53
    matrix calibrate, stage position for sq preset
54
        error ~50%
55
        shift ~25%
56
        num position ~2-3
57
        interval ~1-3e-6 
58
    RCT node calibration
59
        make sure the number of features is between 700 and 1500
60
        take a picture both with and without the grid bar
61
        take a picture both untilted and tilted 
62
    calibrate dose
63
    RCT node setup
64
    re-align presets
65
    re-center objective aperture 
66
67
h2. Running Leginon RCT
68
69
    Start a square:
70
        Square Targeting (if you have an atlas)
71
            Go to "Square Targeting"
72
            Hit the reload button
73
            Select a new square with the green cross
74
            Click the play |> button 
75
        Simulate Square (if you do NOT have an atlas)
76
            Go to "Square"
77
            Hit the simulate target button 
78
    Square centering
79
        When Leginon is finished, a question mark will appear next to the "Square Centering" node
80
        Click on the "Square Centering" node
81
        Select one z-focus position (blue cross) and one imaging position (green position) close to the center of the square
82
        Click the play |> button 
83
    RCT Targeting
84
        When Leginon is finished setting the z-height, a question mark will appear next to the "RCT Targeting" node
85
        Click on the "RCT Targeting" node
86
        Now you select all the holes to image (not too close to the edge) and a single focus position
87
        Optional: obtain a picture of the tilted square
88
            Click on the "Navigation" node
89
            Open up VNC viewer of the tecnai
90
            In the VNC window, set the tilt angle to the desire angle
91
            In the "Navigation" node, collect an image by clicking on the acquire button (camera logo) at the top
92
            Important, in the VNC window, return the tilt angle to zero degrees, otherwise all your images will fail
93
            Now select you targets in the "RCT Targeting" node and view the tilted image in the "Navigation" node. Jump back and forth between the two nodes until you select all holes present in both images. 
94
        When finished, hit the play |> button 
95
    Focusing
96
        If manual focusing is enabled, wait for the manual focus window to pop up
97
        focus the microscope, reset the defocus, and click the red stop button. 
98
    Watch your images in the "RCT" node 
99
100
h2. Nitrogen fill
101
102
Since RCT only focuses once per square, when you fill the dewar, you must refocus to account for changes in the z-height and focus during refill and there is no check for drift calm down so you must manually wait for drift to subside.
103
104
Steps in Nitrogen fill:
105
106
    Pause RCT node
107
    Wait until Leginon is paused
108
    Fill holder and cold trap dewars
109
    Turn off lights
110
    Check if objective aperture is inserted
111
    RESET TIMER
112
    Focus
113
        simulate z-focus
114
        send hole preset to scope
115
        navigate to good focus spot in navigator
116
        simulate focus in the focus node
117
        repeat focus until drift subsides 
118
    If there is an empty hole: you can measure dose or collect bright fields
119
    Declare drift
120
    Hit play button on RCT node
121
    Check images to make sure they are good
122
    Uncheck user verification of targets 
123
124
    This page was last modified 16:45, 16 July 2010.
125 2 Anchi Cheng
126
127
__________________
128
129 3 Anchi Cheng
[[RCT calibration| < RCT application calibration]]