Running "Manual" Application

  1. Leginon/Manual> Open the "Settings" tool and configure the settings:
     
    Important-Use low dose kit option should be set to yes if using the low dose kit on Tecnai. Low Dose Pause Time need to be set to 5 sec or more to complete the mode change.
     
    Recommendations-
    • Acquisition/Camera Configuration- use one of the preconfigured selection for binning and size
       
    • Correct image=yes if dark and bright images of the camera configuration exist.
       
    • Save image to the database=yes.
       
  2. Leginon/manual/Settings> Save and close the settings by pressing "OK" button.
     
  3. Adjust the exposure time in the setting and/or beam size/spot size on the scope to achieve the desired dose and coverage in exposure mode, if low-dose kit is used.
     
  4. Microscope> Search the grid for interested area and focus as normally performed.
     
  5. Microscope> If low dose mode is used, leave TEM in Focus model. Leave the screen at up position if "Main Screen/Up before acquire" is not checked in the settings. No need to put on beam blank.
     
  6. Leginon/manual> Click on the "Acquire" tool to acquire the image.
     
  7. Microscope> If low dose mode is used, Leginon will force the pre-specimen shutter close, switch the scope to exposure mode to acquire the image and then switch back to the original mode.