h1. Setup and Calibration for jeol.cfg
2
3 2 Anchi Cheng
All setup and calibrations are done on the computer that has access to JEOL's External COM through temext.dll.
4 1 Anchi Cheng
5
h2. Copy from pyscope/jeol.cfg.template to pyscope/jeol.cfg
6
7 3 Anchi Cheng
The current template is based on a JEM-3200FSC
8 2 Anchi Cheng
9 3 Anchi Cheng
h2. Modify [tem option] section in pyscope/jeol.cfg according to your scope
10 1 Anchi Cheng
11 3 Anchi Cheng
h3. Using PLA for image shift
12 1 Anchi Cheng
13 3 Anchi Cheng
Image shift is normally achieved with IS1.  However, PLA is a sometimes a better choice, especially if energy filter is involved.
14
15
h3. Lens Series within MAG1 and LOW MAG modes
16
17
There are distinct sets of objective and projection lens settings within MAG1 mode that gives
18
different scale and neutral values.  We assign each of these lens series (LS) numbers, starting at 1 at
19
the lowest mags. Similarly, we also named the lowest LOWMAG mode lens series LM1 and assign a
20
different deflector scale to them.
21
22
The order of magnifications is therefore roughly
23
24
<pre>
25
LM1->the rest of LOWMAG->LS1->LS2->LS3->the rest of MAG1
26
</pre>
27
28
* To simply the mapping of magnification to the projection modes, MAG1 will be used if a magnification
29
can be achieved by either LOWMAG and MAG1*
30
31 5 Anchi Cheng
These four configurations like these therefore define the boundary of each lens series within the mode.
32 3 Anchi Cheng
<pre>
33
LS1_MAG_MAX = 5000
34
LS2_MAG_MAX = 30000
35
LS3_MAG_MAX = 100000
36
LM1_MAG_MAX = 200
37
</pre>
38
39
h2. Run pyscope/scalecalibrator.py to get values for different items in jeol.cfg
40
41 7 Anchi Cheng
With the main viewing screen down do these at the computer with TEM EXT access.
42 4 Anchi Cheng
43 3 Anchi Cheng
# Open pyscope/jeol.cfg with a text editor and leave it open so that you can enter values as you go.
44
# Double click pyscope/scalecalibrator.py to start it.
45 6 Anchi Cheng
# The script will first make the scope go through all magnifications in LOWMAG and MAG1 range to determine valid magnifications.
46 3 Anchi Cheng
# After choosing tem options, enter the magnifications at which to calibrate.  Make sure that you include at least one magnification in each lens series that you will use.
47
# Follow the instruction to shift each coil or mechanism by the amount indicated.  The distance between the major ticks of the linear scale bar on the main viewing screen corresponds to 1 cm at the film camera (i.e., nominal magnification)