Project

General

Profile

Troubles with Targeting » History » Version 13

Anchi Cheng, 10/12/2010 12:20 PM

1 1 Amber Herold
h1. Troubles with Targeting
2
3 12 Anchi Cheng
h2. Quick Links
4
* [[Troubles with Targeting#Targeting-is-consistently-off-in-the-same-direction-and-distance|Targeting is consistently off in the same direction and distance]]
5
* [[Troubles with Targeting#Targeting-is-consistently-off-in-opposite-direction-from-where-you-pick|Targeting is consistently off in opposite direction from where you pick]]
6
* [[Troubles with Targeting#Hole-targeting-is-inconsistently-off-or-goes-bad-in-the-next-run-after-correction|Hole targeting is inconsistently off or goes bad in the next round after correction]]
7 13 Anchi Cheng
* [[Troubles with Targeting#Hole targeting-is-off-more-when-the-image-shift-target-is-away-from-the-center-of-the-image|Hole targeting is off more when the image shift target is away from the center of the image]]
8
* [[Troubles with Targeting#|]]
9 12 Anchi Cheng
* [[Troubles with Targeting#Grid-mosaic-is-too-small-to-see-the-details|Grid mosaic is too small to see the details]]
10
* [[Troubles with Targeting#Grid-mosaic-is-not-patched-together-well|Grid mosaic is not patched together well]]
11
* [[Troubles with Targeting#Grid-mosaic-display-in-Leginon-is-different-from-that-of-the-Web-Viewer|Grid mosaic display in Leginon is different from that of the Web Viewer]]
12 10 Anchi Cheng
13 1 Amber Herold
h2. Grid mosaic is too small to see the details
14
15 3 Eric Hou
h3. Why:
16 4 Eric Hou
17 1 Amber Herold
* The image scale for creating mosaic is set too small.
18
19 3 Eric Hou
h3. Solution:
20 4 Eric Hou
21 3 Eric Hou
#  Leginon/Square Targeting/Mosaic Settings> Change "Scale Image" to a larger
22 1 Amber Herold
size
23 3 Eric Hou
 
24
#  Leginon/Square Targeting/Mosaic Settings> Click "Apply" and then "Create" the
25 1 Amber Herold
mosaic atlas.
26
27
h2. Grid mosaic is not patched together well
28
29 3 Eric Hou
h3. Commonly Why:
30 5 Eric Hou
31 1 Amber Herold
* LM mode U-center defocus not set to 0.
32
33
* LM mode rotation center is off.
34
35
* Bad stage position matrix calibration
36
37 3 Eric Hou
h3. Solutions:
38 1 Amber Herold
39 3 Eric Hou
#  Correct the alignemnt error or redo <link linkend="mcal_st_pos">stage position matrix calibration</link>
40
&nbsp;
41
#  Acquire a new atlas
42
&nbsp;
43
**  Leginon/Grid Targeting/Setting> Enter a new label
44
&nbsp;
45
**  Leginon/Grid Targeting> Click "Calculate Atlas"
46
&nbsp;
47
**  Leginon/Grid Targeting> Click "Submit Targets" (play icon) to begin acquiring the atlas.
48
&nbsp;
49
**  Leginon/Square Targeting> Follow the atlas progress in the display.
50 1 Amber Herold
51 3 Eric Hou
h2. Grid mosaic display in Leginon is different from that of the Web Viewer
52 1 Amber Herold
53 3 Eric Hou
h3. Why:
54 5 Eric Hou
55 3 Eric Hou
* The goniometer does not reach the intended position due to ,most commonly, a pole touch error that temporaily disables the goniometer.
56 1 Amber Herold
57 3 Eric Hou
* The calibration used to move to the target in "Grid" node is not the same as the calibration parameter used to create the mosaic in "Square Targeting" node are not consistent.
58 1 Amber Herold
59 3 Eric Hou
h3. Solution:
60 1 Amber Herold
61 3 Eric Hou
*  Other than fix the goniometer, Targeting on these mosaic can still work if the error does not occur frequently enough to interfere drift management.
62 1 Amber Herold
63 3 Eric Hou
*  Check "Leginon/Grid/Settings/Use ?? to move to target" against "Leginon/Square Targeting/Mosaic Settings/CalibrationParameters". The latter should be changed if the mosaic is already acquire and what seen on the Web is better than what is in Leginon. The mosaic atlas can be recreated by clicking the "Create" button.
64
65 1 Amber Herold
h3. More Explanation:
66 5 Eric Hou
67 1 Amber Herold
*  The web tools and Leginon main program use different mechanism to paste together the atlas.
68
&nbsp;
69 3 Eric Hou
The targets Leginon "Grid Targeting" node send out is an even spaced raster. "Grid" node uses that information and the move type set for the node to calculate the goniometer position it needs to aim for in order to get to where it really wants to go to and acquires the image. It is done this way so that it can handle any complex move type. For example, lets say that the model is p=x^2 so that to get to position p' from 0, the goniometer x needs to be sqrt(p'). "Grid Targeting" send the target as p', "Grid" figure out that x'=sqrt(p') and send the x value to the scope to move.
70 1 Amber Herold
&nbsp;
71 3 Eric Hou
When it needs to display the atlas in "Square targeting", Leginon uses the actual goniometer position of each tile image back calculates the targeted position according to the model. This has the advantage that if there is a problem with the goniometer and the stage does not move to where it needs to (such as after a pole touch error), the tile image will appear at its actual position. In the example above, the image will appear at p' from the image because from the goniometer position x'=sqrt(p'), p=x^2=(sqrt(p'))^2.
72 1 Amber Herold
&nbsp;
73
The webtool, however, pastes the atlas together using the targets saved from "Grid Targeting". It does not care what you have set in "Square Targeting" node. Therefore, it will always look right, unless of course the stage did not move to the target when it acquired the image.
74
75 3 Eric Hou
h2. Targeting is consistently off in the same direction and distance
76 1 Amber Herold
77 3 Eric Hou
Commonly Why: Image shifts in "hl", "sq", or "gr"presets are not aligned with the presets at higher mags. The low mag "sq" preset is especially prone to this problem.
78 1 Amber Herold
79 8 Anchi Cheng
*Important: When the targeted image is off, it normally means the image shift of the preset where the target comes from is off. That is, if hole images appear off, it is not the image shift of "hl" preset that needs correcting but that of "sq" preset.*
80 1 Amber Herold
81
Solution:
82
83 3 Eric Hou
*  Pause before next hole or square by clicking Hole/Toolbar>Pause button or Square/Toolbar>Pause button, respectively. Which one to pause depends on what is the next target.
84 1 Amber Herold
85
*  Center the grid at a location that can be recognized by either:
86 3 Eric Hou
&nbsp;
87
##  Move to the alignment position:
88
&nbsp;
89
(It may be necessary to adjust the position at the scope after this since the goniometer can not move back to a position as accurately as we need it).
90
&nbsp;
91
***  Leginon/Navigation/Stage Location> Choose the name for the stored alignment position ("align" if you have been following the setup check list). and send it "To Scope"
92
&nbsp;
93
##  scope> move the specimen to a position on the same square, either on a recognizable object or burn a good hole through ice as preparation for lower mag preset image shift correction
94 1 Amber Herold
95 9 Anchi Cheng
Perform [[MSI_set-up_in_more_details#Medium-mag-preset-image-and_beam-shift-refinement|Preset Image Shift Alignment]]
96 1 Amber Herold
97 9 Anchi Cheng
Still not working: You may have an invalid combination of settings such as using [[Iterative Stage Movement]] while [[Queuing Option|queuing up]] targets on the images so obtained.
98 1 Amber Herold
99
Solution:
100
101 9 Anchi Cheng
*  Resolve the conflict.
102 1 Amber Herold
103 3 Eric Hou
h2. Targeting is consistently off in opposite direction from where you pick
104 1 Amber Herold
105
Commonly Why: Camera configuration is different in calibration and now.
106
107 3 Eric Hou
#  All your targeting would be wrong. "Abort" target processing in the Target Watcher of "squares" and "holes".
108
&nbsp;
109
#  If you use the Tietz camera: scope/Leginon/Instrument/Camera> check your camera configuration. mirror should be ['vertical']. Change it if not, and "Set" the configuration.
110
&nbsp;
111
#  If you use the Gatan camera: scope/Digital Micrograph> check your camera configuration.
112
&nbsp;
113
#  Leginon/Presets Manager> send "sq" or "grid" "To Scope"
114
&nbsp;
115
#  Leginon/Navigation> "Acquire" an image with "Use this configuration" Unchecked in the Camera Configuration.
116
&nbsp;
117
#  Leginon/Navigation> "Navigate" by "Image Shift", "Beam Shift", or "Stage Position" to center on a reference object or burn mark made in the high mag presets. This is to check if it behaves correctly now.
118
&nbsp;
119
#  Select targets again, or, if atlas is wrong, republish the target list to make the atlas.
120 1 Amber Herold
121 3 Eric Hou
h2. Hole targeting is inconsistently off or goes bad in the next run after correction
122 1 Amber Herold
123 3 Eric Hou
Commonly Why: Not well understood but may be related to microscopy alignment and settings such as normalization of lenses. up to 1.5 um error has been observed even with a reasonable calibration. The following solutions only works some times.
124 1 Amber Herold
125 3 Eric Hou
#  Leginon/Drift Manager>"Declare Drift" to force correction for possible drift.
126
&nbsp;
127 6 Eric Hou
#  Repeat image shift correction alignment [[Troubles with Targeting#Targeting_is_consistently_off_in_the_same_direction_and_distance|as in the case when targeting is consistently off]].
128 3 Eric Hou
&nbsp;
129
#  Ignore the first targetted '"hole" in the "squares" image. The second and on are more consistent when there is a hysteresis problem.
130
&nbsp;
131 7 Eric Hou
#  Repeat [[Modeled_Stage_Position_calibration|"Mag Only" Modeled Stage Calibration]] at the "square" preset.
132 1 Amber Herold
133 3 Eric Hou
h2. Hole targeting is off more when the image shift target is away from the center of the image
134 1 Amber Herold
135 3 Eric Hou
Commonly Why: Not well understood but may be related to microscopy image shift causes different physical movement at different magnifications since Leginon uses the image shift calibration at the lower mag where the target comes from to determine the required image shift at the higher mag.
136 1 Amber Herold
137
No current solution
138
139 3 Eric Hou
h2. Automatic hole finding fails to pick up good ice thickness after set up
140 1 Amber Herold
141
Commonly Why:
142 3 Eric Hou
143 1 Amber Herold
* Electron beam intensity has changed
144
145
* Blob threshold is not inclusive enough to tolerate variations
146
147
* Ice contaimination
148
149
Solutions
150
151 3 Eric Hou
* Monitor electron beam with a saved location with an empty hole in navigator and check the recorded intensity periodically and adjust zero ice thickness accordingly.
152 1 Amber Herold
153
* Reduce Blob threshold in Alias nodes of the Hole Finder.
154
155
* Make sure the anticontamination cryo box is properly inserted on the scope.
156
157
* Remove the film camera from the scope if you don't need film.
158 2 Amber Herold
159 3 Eric Hou
h2. No broken grid square for dose measurement and ZLP alignment (MSI-Tomography)
160 2 Amber Herold
161
Solution
162
163 3 Eric Hou
* Burn a hole clear of ice and pick it as the reference target in "Hole Targeting" or "Tomography Targeting"
164 2 Amber Herold
165
______
166
167
[[Troubles with Imaging - Troubleshooting|< Troubles with Imaging]] | [[Troubles with Focusing and Drift Check|Troubles with Focusing and Drift Check >]]
168
169
______