Actions
Bug #1569
closeddefocus of the second of two autofocus sequence with the same preset and defocus adjustment is not set
Status:
Closed
Priority:
Normal
Assignee:
Category:
Leginon subsystem
Target version:
Start date:
02/29/2012
Due date:
% Done:
0%
Estimated time:
Affected Version:
Appion/Leginon 2.1.0
Show in known bugs:
No
Workaround:
Description
Since the preset is not sent again to save time, the final defocus of the first (O defocus) is used in the second sequence. This makes the second one often fails and therefore set to eucentric focus and the first effort is not taking effect.