defocus of the second of two autofocus sequence with the same preset and defocus adjustment is not set

Added by Anchi Cheng over 12 years ago. Updated over 12 years ago.

Status:
Closed
Priority:
Normal
Assignee:
Category:
Leginon subsystem
Start date:
02/29/2012
Due date:
% Done:

0%

Estimated time:
Affected Version:
Appion/Leginon 2.1.0
Show in known bugs:
No
Workaround:

Description

Since the preset is not sent again to save time, the final defocus of the first (O defocus) is used in the second sequence. This makes the second one often fails and therefore set to eucentric focus and the first effort is not taking effect.


Related issues 1 (0 open1 closed)

Has duplicate Leginon - Bug #562: Focus sequence problemClosedJim Pulokas05/25/2010