h1. Full Protocol on a F30 with an energy filter
2
3
4
5
h2.  Full Protocol on a F30 with an energy filter
6
7
8
9
This is edited from a complete protocol written by Lu Gan (Jensen Lab, Caltech) for users
10
who have not used general MSI applications. It is modified to reflect the current Leginon
11
version and a current and graphic-rich protocol by Jian Shi (contact him at
12
jianshi@caltech.edu for a copy) of the same group.
13
14
15
16
h2. Notes
17
18
19
20
*  Make sure EFTEM is always on
21
22
23
*  If presets manager has error, disable then enable "Filtered" box in
24
Tecnai/TUI/GIF> menu. (TUI: Tecnai User Interface)
25
26
27
*  Leginon re-aligns ZLP, using a "reference" hole, after each tilt series.
28
29
30
* Pause Leginon (for filling dewars, etc.): wait until tilt series is almost done,
31
then press "PAUSE" in FOCUS node. Leginon will pause at start of next tilt
32
series.
33
34
35
* Overall scheme: Take grid atlas. At reference position, do cross-over correction,
36
align presets, center objective aperture. Select and preview targets. Prepare data
37
collection (gr and sq presets = skip sequence). Take dose in tomography setting. Let
38
objective lens equilibrate. Align ZLP, insert objective aperture, fire away!
39
40
41
42
43
44
45
46
h2. Initial setup
47
48
49
50
*  Tecnai/GIF> Tune GIF with scope column empty
51
52
53
*  Tecnai/TUI/Stage> Set Z = 0.
54
55
56
*  Tecnai/TUI> Turn on Intensity Zoom
57
58
59
*  Tecnai> Remove objective aperture.
60
61
62
*  Quit UCSF Tomo, TEM Auto-functions, Low-Dose kit.
63
64
65
66
67
68
69
70
h2. Start Leginon
71
72
73
74
*  Launch Leginon
75
76
77
*  Choose either to start a new session or continue an old one, then launch Multiscale
78
Tomography from Applications menu.
79
80
81
82
83
84
85
86
h2. Presets setup with empty column
87
88
89
90
*  Leginon/Presets Manager> Set up all the presets except align image shifts. If
91
starting new session, import presets from previous session. If you want to preview
92
images at high mag, make or copy a preview called "preview," which should be set to 3x3
93
to 4x4 binning for greater contrast.
94
95
96
97
98
99
100
101
h2. Flat fielding and dose calibration
102
103
104
105
*  Tecnai/TUI/EFTEM> Make sure that "filter" is on.
106
107
108
*  Leginon/Presets Manager> tomo preset to scope.
109
110
111
*  Tecnai/TUI/GIF> Go back to hole and re-align ZLP (and tune GIF if it's been ~24
112
hours since last
113
114
115
*  Leginon/Correction/Settings> set camera configuration to match that of the
116
preset. Choose to use 2 correction channels.
117
118
119
*  Leginon/Correction> acquire dark current, then acquire bright image, then
120
acquire corrected image to make sure "flat-fielding" worked.
121
122
123
*  Leginon/Presets manager> redo dose measurement.
124
125
126
*  Leginon/Tomography> recheck dose.
127
128
129
*  Leginon/Correction> repeat 4 and 5 for preset fa and hl if they are of different
130
camera binning.
131
132
133
134
135
136
137
138
h2. Presets setup and grid atlas
139
140
141
142
*  Insert Grid.
143
144
145
*  Make sure you're near eucentric height (z = 0 is close enough for Leginon, unless
146
your grid is seriously bent).
147
148
149
*  You should know from experience whether the imported "gr" preset aligns reasonable
150
well to higher mags so that you can reach your intended grid square, however inaccurate,
151
without alignment. If this is true, you can follow 4 and 5 immediately. If not, you will
152
need to manually move to a test sacrificial square at Tecnai and acquire the grid atlas
153
after presets alignment.
154
155
156
*  Leginon/Grid Targeting> Calculate and then publish a grid atlas using the "Grid
157
Targeting" node. Do NOT pick targets in "Square" node during atlas assembly.
158
159
160
*  Leginon/Square Targeting> Pick an broken square on the grid atlas and press the
161
play icon, then save position on Tecnai/TUI/Stage> or in Leginon/Navigation/Stage
162
Locations>. Repeat for a sacrificial square. The broken square is used to manually
163
align the ZLP; the sacrificial square is used to align presets, objective lens, and to
164
record test images.
165
166
167
168
169
170
171
172
h2. Presets alignment
173
174
175
Note: steps 1& 2 are only necessary for really bad grids (you'll know).
176
177
178
*  Tecnai/TUI> stage: go to the sacrificial square. Turn on wobbler
179
180
181
*  Tecnai/DM> do auto eucentric height and auto focus.
182
183
184
*  Leginon/Presets Manager> hl preset to scope.
185
186
187
*  Tecnai/TUI> move the stage to a feature recongnizable across the presets.
188
189
190
*  Leginon/Presets Manager> tomo preset to scope.
191
192
193
*  Tecnai/TUI> recenter the feature accurately.
194
195
196
*  Leginon/Presets Manager> Follow <link linkend="MSI_image_shift_sq">image shift
197
alignment procedure</link> in general MSI set-up using tomo preset as the
198
reference.
199
200
201
202
203
204
205
206
h2. Objective aperture centering
207
208
209
210
*  Leginon/Presets Manager> tomography presets to scope.
211
212
213
*  Tecnai/TUI/motorized apertures> Put in objective aperture.
214
215
216
*  Tecnai> Go into diffraction mode.
217
218
219
*  Tecnai> Center image using diffraction shift-X&amp;Y.
220
221
222
*  Tecnai/TUI/motorized apertures> Click center in motorized apertures menu.
223
224
225
*  Tecnai/TUI/motorized apertures> Center objective aperture, then remove objective
226
aperture.
227
228
229
230
231
232
233
234
h2. Target acquisition
235
236
237
238
* Leginon/Grid Targeting> Remove objective aperture, Rename and calculate a new
239
grid atlas, then publish to make a new atlas, and then Reinsert objective
240
aperture.
241
242
243
***Note: You only need to take this (second) atlas if the presets - especially
244
square and hole - are poorly aligned, which you'll find out from the previous
245
alignments. If the presets were already aligned, then you can use the original grid
246
atlas. ***
247
248
249
*  Leginon/Square Targeting> Click on the purple crosshair icon (reference point)
250
and click on a broken square, which will be used for ZLP alignment and dosage.
251
252
253
*  Leginon/Square Targeting> Pick squares that look good, then press Play button
254
!http://emg.nysbc.org/software/leginon/images/icons/play.png!.
255
256
257
*  Leginon/Hole Targeting> While square targets are getting collected, you can
258
start picking targets in the hole targeting menu. Pick a bunch of holes (but not
259
adjacent ones), then pick a focus point for rough Z- focus. To advance to the next
260
square, press the play button but NOT Queue- play!http://emg.nysbc.org/software/leginon/images/icons/send_queue_out.png!.
261
262
263
*  Leginon/Hole Targeting> After the hole targets are all selected, insert the
264
objective aperture and press Queue-play.
265
266
267
*  Leginon/Tomography Targeting> DON'T PRESS Queue-play until ready for final
268
tomography acquisition.
269
270
271
*  Leginon/Tomography Targeting>Remember to pick a FOCUS spot for each hole and
272
tomography target!
273
274
275
Tip: you can mouse over the target and get an estimate of the average intensity in
276
the popup menu, which lists X,Y,Cnts. This feature lets you rapidly screen for (1)
277
squares with the thinnest ice and (2) the thinnest targets.
278
279
280
*  Leginon/Tomography Targeting> To preview target, use preview crosshair to pick
281
the target:
282
* SET defocus in preview preset to -1.2 e-06 m and adjust the exposure time (in
283
milliseconds) such that the dose is ~ 0.2 - 0.5 e-/Angstrum^2 - VERY
284
IMPORTANT!
285
286
287
* If you pick a preview target and press Play button, it will be processed
288
immediately. Therefore, don't do it until all hl images are taken.
289
290
291
* After you have inspected the preview image, you can come back to Tomography
292
Targeting node to pick final targets on the same image. Alternatively, you can
293
pick the targets first with the preview target and then remove the unwanted ones
294
after previewing.
295
296
297
298
299
300
*  Leginon/Tomography> check data collection parameters: dose, tilt range, tilt
301
interval. See <link linkend="Tomo_Intensity">notes on image intensity recorded through
302
tomography</link>. (If you want -60 to 60 tilts, you should ender -61 to 61 since the
303
limiting algorithm is "smaller than" not "smaller than or equal to".
304
305
306
*  Leginon/Presets> setup data collection parameters for focus &amp; tomography:
307
mag, defocus, INT (INT for focus should be greater than for tomo).
308
309
310
311
312
313
314
315
h2. Final checklist: Getting ready to collect
316
317
318
319
*  Leginon/Presets manager> tomo preset to scope.
320
321
322
*  Tecnai/TUI/Stage> Go to reference point.
323
324
325
*  Leginon/Tomography node> re-check dose and re-center beam.
326
327
328
*  Leginon/Exposure targeting> press queue-play. Atlas does NOT need to be
329
reloaded.
330
331
332
333
334
335
336
337
338
339