Full Protocol on a F30 with an energy filter » History » Version 1
Amber Herold, 06/21/2010 12:02 PM
1 | 1 | Amber Herold | h1. Full Protocol on a F30 with an energy filter |
---|---|---|---|
2 | |||
3 | |||
4 | |||
5 | h2. Full Protocol on a F30 with an energy filter |
||
6 | |||
7 | |||
8 | |||
9 | This is edited from a complete protocol written by Lu Gan (Jensen Lab, Caltech) for users |
||
10 | who have not used general MSI applications. It is modified to reflect the current Leginon |
||
11 | version and a current and graphic-rich protocol by Jian Shi (contact him at |
||
12 | jianshi@caltech.edu for a copy) of the same group. |
||
13 | |||
14 | |||
15 | |||
16 | h2. Notes |
||
17 | |||
18 | |||
19 | |||
20 | * Make sure EFTEM is always on |
||
21 | |||
22 | |||
23 | * If presets manager has error, disable then enable "Filtered" box in |
||
24 | Tecnai/TUI/GIF> menu. (TUI: Tecnai User Interface) |
||
25 | |||
26 | |||
27 | * Leginon re-aligns ZLP, using a "reference" hole, after each tilt series. |
||
28 | |||
29 | |||
30 | * Pause Leginon (for filling dewars, etc.): wait until tilt series is almost done, |
||
31 | then press "PAUSE" in FOCUS node. Leginon will pause at start of next tilt |
||
32 | series. |
||
33 | |||
34 | |||
35 | * Overall scheme: Take grid atlas. At reference position, do cross-over correction, |
||
36 | align presets, center objective aperture. Select and preview targets. Prepare data |
||
37 | collection (gr and sq presets = skip sequence). Take dose in tomography setting. Let |
||
38 | objective lens equilibrate. Align ZLP, insert objective aperture, fire away! |
||
39 | |||
40 | |||
41 | |||
42 | |||
43 | |||
44 | |||
45 | |||
46 | h2. Initial setup |
||
47 | |||
48 | |||
49 | |||
50 | * Tecnai/GIF> Tune GIF with scope column empty |
||
51 | |||
52 | |||
53 | * Tecnai/TUI/Stage> Set Z = 0. |
||
54 | |||
55 | |||
56 | * Tecnai/TUI> Turn on Intensity Zoom |
||
57 | |||
58 | |||
59 | * Tecnai> Remove objective aperture. |
||
60 | |||
61 | |||
62 | * Quit UCSF Tomo, TEM Auto-functions, Low-Dose kit. |
||
63 | |||
64 | |||
65 | |||
66 | |||
67 | |||
68 | |||
69 | |||
70 | h2. Start Leginon |
||
71 | |||
72 | |||
73 | |||
74 | * Launch Leginon |
||
75 | |||
76 | |||
77 | * Choose either to start a new session or continue an old one, then launch Multiscale |
||
78 | Tomography from Applications menu. |
||
79 | |||
80 | |||
81 | |||
82 | |||
83 | |||
84 | |||
85 | |||
86 | h2. Presets setup with empty column |
||
87 | |||
88 | |||
89 | |||
90 | * Leginon/Presets Manager> Set up all the presets except align image shifts. If |
||
91 | starting new session, import presets from previous session. If you want to preview |
||
92 | images at high mag, make or copy a preview called "preview," which should be set to 3x3 |
||
93 | to 4x4 binning for greater contrast. |
||
94 | |||
95 | |||
96 | |||
97 | |||
98 | |||
99 | |||
100 | |||
101 | h2. Flat fielding and dose calibration |
||
102 | |||
103 | |||
104 | |||
105 | * Tecnai/TUI/EFTEM> Make sure that "filter" is on. |
||
106 | |||
107 | |||
108 | * Leginon/Presets Manager> tomo preset to scope. |
||
109 | |||
110 | |||
111 | * Tecnai/TUI/GIF> Go back to hole and re-align ZLP (and tune GIF if it's been ~24 |
||
112 | hours since last |
||
113 | |||
114 | |||
115 | * Leginon/Correction/Settings> set camera configuration to match that of the |
||
116 | preset. Choose to use 2 correction channels. |
||
117 | |||
118 | |||
119 | * Leginon/Correction> acquire dark current, then acquire bright image, then |
||
120 | acquire corrected image to make sure "flat-fielding" worked. |
||
121 | |||
122 | |||
123 | * Leginon/Presets manager> redo dose measurement. |
||
124 | |||
125 | |||
126 | * Leginon/Tomography> recheck dose. |
||
127 | |||
128 | |||
129 | * Leginon/Correction> repeat 4 and 5 for preset fa and hl if they are of different |
||
130 | camera binning. |
||
131 | |||
132 | |||
133 | |||
134 | |||
135 | |||
136 | |||
137 | |||
138 | h2. Presets setup and grid atlas |
||
139 | |||
140 | |||
141 | |||
142 | * Insert Grid. |
||
143 | |||
144 | |||
145 | * Make sure you're near eucentric height (z = 0 is close enough for Leginon, unless |
||
146 | your grid is seriously bent). |
||
147 | |||
148 | |||
149 | * You should know from experience whether the imported "gr" preset aligns reasonable |
||
150 | well to higher mags so that you can reach your intended grid square, however inaccurate, |
||
151 | without alignment. If this is true, you can follow 4 and 5 immediately. If not, you will |
||
152 | need to manually move to a test sacrificial square at Tecnai and acquire the grid atlas |
||
153 | after presets alignment. |
||
154 | |||
155 | |||
156 | * Leginon/Grid Targeting> Calculate and then publish a grid atlas using the "Grid |
||
157 | Targeting" node. Do NOT pick targets in "Square" node during atlas assembly. |
||
158 | |||
159 | |||
160 | * Leginon/Square Targeting> Pick an broken square on the grid atlas and press the |
||
161 | play icon, then save position on Tecnai/TUI/Stage> or in Leginon/Navigation/Stage |
||
162 | Locations>. Repeat for a sacrificial square. The broken square is used to manually |
||
163 | align the ZLP; the sacrificial square is used to align presets, objective lens, and to |
||
164 | record test images. |
||
165 | |||
166 | |||
167 | |||
168 | |||
169 | |||
170 | |||
171 | |||
172 | h2. Presets alignment |
||
173 | |||
174 | |||
175 | Note: steps 1& 2 are only necessary for really bad grids (you'll know). |
||
176 | |||
177 | |||
178 | * Tecnai/TUI> stage: go to the sacrificial square. Turn on wobbler |
||
179 | |||
180 | |||
181 | * Tecnai/DM> do auto eucentric height and auto focus. |
||
182 | |||
183 | |||
184 | * Leginon/Presets Manager> hl preset to scope. |
||
185 | |||
186 | |||
187 | * Tecnai/TUI> move the stage to a feature recongnizable across the presets. |
||
188 | |||
189 | |||
190 | * Leginon/Presets Manager> tomo preset to scope. |
||
191 | |||
192 | |||
193 | * Tecnai/TUI> recenter the feature accurately. |
||
194 | |||
195 | |||
196 | * Leginon/Presets Manager> Follow <link linkend="MSI_image_shift_sq">image shift |
||
197 | alignment procedure</link> in general MSI set-up using tomo preset as the |
||
198 | reference. |
||
199 | |||
200 | |||
201 | |||
202 | |||
203 | |||
204 | |||
205 | |||
206 | h2. Objective aperture centering |
||
207 | |||
208 | |||
209 | |||
210 | * Leginon/Presets Manager> tomography presets to scope. |
||
211 | |||
212 | |||
213 | * Tecnai/TUI/motorized apertures> Put in objective aperture. |
||
214 | |||
215 | |||
216 | * Tecnai> Go into diffraction mode. |
||
217 | |||
218 | |||
219 | * Tecnai> Center image using diffraction shift-X&Y. |
||
220 | |||
221 | |||
222 | * Tecnai/TUI/motorized apertures> Click center in motorized apertures menu. |
||
223 | |||
224 | |||
225 | * Tecnai/TUI/motorized apertures> Center objective aperture, then remove objective |
||
226 | aperture. |
||
227 | |||
228 | |||
229 | |||
230 | |||
231 | |||
232 | |||
233 | |||
234 | h2. Target acquisition |
||
235 | |||
236 | |||
237 | |||
238 | * Leginon/Grid Targeting> Remove objective aperture, Rename and calculate a new |
||
239 | grid atlas, then publish to make a new atlas, and then Reinsert objective |
||
240 | aperture. |
||
241 | |||
242 | |||
243 | ***Note: You only need to take this (second) atlas if the presets - especially |
||
244 | square and hole - are poorly aligned, which you'll find out from the previous |
||
245 | alignments. If the presets were already aligned, then you can use the original grid |
||
246 | atlas. *** |
||
247 | |||
248 | |||
249 | * Leginon/Square Targeting> Click on the purple crosshair icon (reference point) |
||
250 | and click on a broken square, which will be used for ZLP alignment and dosage. |
||
251 | |||
252 | |||
253 | * Leginon/Square Targeting> Pick squares that look good, then press Play button |
||
254 | !http://emg.nysbc.org/software/leginon/images/icons/play.png!. |
||
255 | |||
256 | |||
257 | * Leginon/Hole Targeting> While square targets are getting collected, you can |
||
258 | start picking targets in the hole targeting menu. Pick a bunch of holes (but not |
||
259 | adjacent ones), then pick a focus point for rough Z- focus. To advance to the next |
||
260 | square, press the play button but NOT Queue- play!http://emg.nysbc.org/software/leginon/images/icons/send_queue_out.png!. |
||
261 | |||
262 | |||
263 | * Leginon/Hole Targeting> After the hole targets are all selected, insert the |
||
264 | objective aperture and press Queue-play. |
||
265 | |||
266 | |||
267 | * Leginon/Tomography Targeting> DON'T PRESS Queue-play until ready for final |
||
268 | tomography acquisition. |
||
269 | |||
270 | |||
271 | * Leginon/Tomography Targeting>Remember to pick a FOCUS spot for each hole and |
||
272 | tomography target! |
||
273 | |||
274 | |||
275 | Tip: you can mouse over the target and get an estimate of the average intensity in |
||
276 | the popup menu, which lists X,Y,Cnts. This feature lets you rapidly screen for (1) |
||
277 | squares with the thinnest ice and (2) the thinnest targets. |
||
278 | |||
279 | |||
280 | * Leginon/Tomography Targeting> To preview target, use preview crosshair to pick |
||
281 | the target: |
||
282 | * SET defocus in preview preset to -1.2 e-06 m and adjust the exposure time (in |
||
283 | milliseconds) such that the dose is ~ 0.2 - 0.5 e-/Angstrum^2 - VERY |
||
284 | IMPORTANT! |
||
285 | |||
286 | |||
287 | * If you pick a preview target and press Play button, it will be processed |
||
288 | immediately. Therefore, don't do it until all hl images are taken. |
||
289 | |||
290 | |||
291 | * After you have inspected the preview image, you can come back to Tomography |
||
292 | Targeting node to pick final targets on the same image. Alternatively, you can |
||
293 | pick the targets first with the preview target and then remove the unwanted ones |
||
294 | after previewing. |
||
295 | |||
296 | |||
297 | |||
298 | |||
299 | |||
300 | * Leginon/Tomography> check data collection parameters: dose, tilt range, tilt |
||
301 | interval. See <link linkend="Tomo_Intensity">notes on image intensity recorded through |
||
302 | tomography</link>. (If you want -60 to 60 tilts, you should ender -61 to 61 since the |
||
303 | limiting algorithm is "smaller than" not "smaller than or equal to". |
||
304 | |||
305 | |||
306 | * Leginon/Presets> setup data collection parameters for focus & tomography: |
||
307 | mag, defocus, INT (INT for focus should be greater than for tomo). |
||
308 | |||
309 | |||
310 | |||
311 | |||
312 | |||
313 | |||
314 | |||
315 | h2. Final checklist: Getting ready to collect |
||
316 | |||
317 | |||
318 | |||
319 | * Leginon/Presets manager> tomo preset to scope. |
||
320 | |||
321 | |||
322 | * Tecnai/TUI/Stage> Go to reference point. |
||
323 | |||
324 | |||
325 | * Leginon/Tomography node> re-check dose and re-center beam. |
||
326 | |||
327 | |||
328 | * Leginon/Exposure targeting> press queue-play. Atlas does NOT need to be |
||
329 | reloaded. |
||
330 | |||
331 | |||
332 | |||
333 | |||
334 | |||
335 | |||
336 | |||
337 | |||