Full Protocol on a F30 with an energy filter » History » Version 5
Amber Herold, 06/21/2010 12:32 PM
1 | 1 | Amber Herold | h1. Full Protocol on a F30 with an energy filter |
---|---|---|---|
2 | |||
3 | |||
4 | |||
5 | |||
6 | |||
7 | This is edited from a complete protocol written by Lu Gan (Jensen Lab, Caltech) for users |
||
8 | who have not used general MSI applications. It is modified to reflect the current Leginon |
||
9 | version and a current and graphic-rich protocol by Jian Shi (contact him at |
||
10 | jianshi@caltech.edu for a copy) of the same group. |
||
11 | |||
12 | |||
13 | |||
14 | h2. Notes |
||
15 | |||
16 | |||
17 | |||
18 | * Make sure EFTEM is always on |
||
19 | |||
20 | |||
21 | * If presets manager has error, disable then enable "Filtered" box in |
||
22 | Tecnai/TUI/GIF> menu. (TUI: Tecnai User Interface) |
||
23 | |||
24 | |||
25 | * Leginon re-aligns ZLP, using a "reference" hole, after each tilt series. |
||
26 | |||
27 | |||
28 | * Pause Leginon (for filling dewars, etc.): wait until tilt series is almost done, |
||
29 | then press "PAUSE" in FOCUS node. Leginon will pause at start of next tilt |
||
30 | series. |
||
31 | |||
32 | |||
33 | * Overall scheme: Take grid atlas. At reference position, do cross-over correction, |
||
34 | align presets, center objective aperture. Select and preview targets. Prepare data |
||
35 | collection (gr and sq presets = skip sequence). Take dose in tomography setting. Let |
||
36 | objective lens equilibrate. Align ZLP, insert objective aperture, fire away! |
||
37 | |||
38 | |||
39 | |||
40 | |||
41 | |||
42 | |||
43 | |||
44 | h2. Initial setup |
||
45 | |||
46 | |||
47 | |||
48 | 2 | Amber Herold | # Tecnai/GIF> Tune GIF with scope column empty |
49 | # Tecnai/TUI/Stage> Set Z = 0. |
||
50 | # Tecnai/TUI> Turn on Intensity Zoom |
||
51 | # Tecnai> Remove objective aperture. |
||
52 | # Quit UCSF Tomo, TEM Auto-functions, Low-Dose kit. |
||
53 | 1 | Amber Herold | |
54 | |||
55 | |||
56 | |||
57 | |||
58 | |||
59 | |||
60 | h2. Start Leginon |
||
61 | |||
62 | 2 | Amber Herold | # Launch Leginon |
63 | # Choose either to start a new session or continue an old one, then launch Multiscale |
||
64 | 1 | Amber Herold | Tomography from Applications menu. |
65 | |||
66 | |||
67 | |||
68 | |||
69 | |||
70 | |||
71 | |||
72 | h2. Presets setup with empty column |
||
73 | |||
74 | |||
75 | |||
76 | 2 | Amber Herold | # Leginon/Presets Manager> Set up all the presets except align image shifts. If |
77 | 1 | Amber Herold | starting new session, import presets from previous session. If you want to preview |
78 | images at high mag, make or copy a preview called "preview," which should be set to 3x3 |
||
79 | to 4x4 binning for greater contrast. |
||
80 | |||
81 | |||
82 | |||
83 | |||
84 | |||
85 | |||
86 | |||
87 | h2. Flat fielding and dose calibration |
||
88 | |||
89 | |||
90 | |||
91 | 2 | Amber Herold | # Tecnai/TUI/EFTEM> Make sure that "filter" is on. |
92 | # Leginon/Presets Manager> tomo preset to scope. |
||
93 | # Tecnai/TUI/GIF> Go back to hole and re-align ZLP (and tune GIF if it's been ~24 |
||
94 | 1 | Amber Herold | hours since last |
95 | 2 | Amber Herold | # Leginon/Correction/Settings> set camera configuration to match that of the |
96 | 1 | Amber Herold | preset. Choose to use 2 correction channels. |
97 | 2 | Amber Herold | # Leginon/Correction> acquire dark current, then acquire bright image, then |
98 | 1 | Amber Herold | acquire corrected image to make sure "flat-fielding" worked. |
99 | 2 | Amber Herold | # Leginon/Presets manager> redo dose measurement. |
100 | # Leginon/Tomography> recheck dose. |
||
101 | # Leginon/Correction> repeat 4 and 5 for preset fa and hl if they are of different |
||
102 | 1 | Amber Herold | camera binning. |
103 | |||
104 | |||
105 | |||
106 | |||
107 | |||
108 | |||
109 | |||
110 | h2. Presets setup and grid atlas |
||
111 | |||
112 | |||
113 | |||
114 | 2 | Amber Herold | # Insert Grid. |
115 | # Make sure you're near eucentric height (z = 0 is close enough for Leginon, unless |
||
116 | 1 | Amber Herold | your grid is seriously bent). |
117 | 2 | Amber Herold | # You should know from experience whether the imported "gr" preset aligns reasonable |
118 | 1 | Amber Herold | well to higher mags so that you can reach your intended grid square, however inaccurate, |
119 | without alignment. If this is true, you can follow 4 and 5 immediately. If not, you will |
||
120 | need to manually move to a test sacrificial square at Tecnai and acquire the grid atlas |
||
121 | after presets alignment. |
||
122 | 2 | Amber Herold | # Leginon/Grid Targeting> Calculate and then publish a grid atlas using the "Grid |
123 | 1 | Amber Herold | Targeting" node. Do NOT pick targets in "Square" node during atlas assembly. |
124 | 2 | Amber Herold | # Leginon/Square Targeting> Pick an broken square on the grid atlas and press the |
125 | 1 | Amber Herold | play icon, then save position on Tecnai/TUI/Stage> or in Leginon/Navigation/Stage |
126 | Locations>. Repeat for a sacrificial square. The broken square is used to manually |
||
127 | align the ZLP; the sacrificial square is used to align presets, objective lens, and to |
||
128 | record test images. |
||
129 | |||
130 | |||
131 | |||
132 | |||
133 | |||
134 | |||
135 | |||
136 | h2. Presets alignment |
||
137 | |||
138 | |||
139 | 2 | Amber Herold | Note: steps 1 and 2 are only necessary for really bad grids (you'll know). |
140 | 1 | Amber Herold | |
141 | |||
142 | 2 | Amber Herold | # Tecnai/TUI> stage: go to the sacrificial square. Turn on wobbler |
143 | # Tecnai/DM> do auto eucentric height and auto focus. |
||
144 | # Leginon/Presets Manager> hl preset to scope. |
||
145 | # Tecnai/TUI> move the stage to a feature recongnizable across the presets. |
||
146 | # Leginon/Presets Manager> tomo preset to scope. |
||
147 | # Tecnai/TUI> recenter the feature accurately. |
||
148 | 4 | Amber Herold | # Leginon/Presets Manager> Follow [[MSI_set_up_in_more_details#Medium-mag-preset-image-and-beam-shift-refinement|image shift |
149 | 2 | Amber Herold | alignment procedure]] in general MSI set-up using tomo preset as the |
150 | 1 | Amber Herold | reference. |
151 | |||
152 | |||
153 | |||
154 | |||
155 | |||
156 | |||
157 | |||
158 | h2. Objective aperture centering |
||
159 | |||
160 | |||
161 | |||
162 | 5 | Amber Herold | # Leginon/Presets Manager> tomography presets to scope. |
163 | # Tecnai/TUI/motorized apertures> Put in objective aperture. |
||
164 | # Tecnai> Go into diffraction mode. |
||
165 | # Tecnai> Center image using diffraction shift-X&Y. |
||
166 | # Tecnai/TUI/motorized apertures> Click center in motorized apertures menu. |
||
167 | # Tecnai/TUI/motorized apertures> Center objective aperture, then remove objective |
||
168 | 1 | Amber Herold | aperture. |
169 | |||
170 | |||
171 | |||
172 | |||
173 | |||
174 | |||
175 | |||
176 | h2. Target acquisition |
||
177 | |||
178 | |||
179 | |||
180 | 5 | Amber Herold | 1. Leginon/Grid Targeting> Remove objective aperture, Rename and calculate a new |
181 | 1 | Amber Herold | grid atlas, then publish to make a new atlas, and then Reinsert objective |
182 | aperture. |
||
183 | |||
184 | |||
185 | ***Note: You only need to take this (second) atlas if the presets - especially |
||
186 | square and hole - are poorly aligned, which you'll find out from the previous |
||
187 | alignments. If the presets were already aligned, then you can use the original grid |
||
188 | atlas. *** |
||
189 | |||
190 | |||
191 | 5 | Amber Herold | 2. Leginon/Square Targeting> Click on the purple crosshair icon (reference point) |
192 | 1 | Amber Herold | and click on a broken square, which will be used for ZLP alignment and dosage. |
193 | |||
194 | |||
195 | 5 | Amber Herold | 3. Leginon/Square Targeting> Pick squares that look good, then press Play button !http://emg.nysbc.org/software/leginon/images/icons/play.png!. |
196 | 1 | Amber Herold | |
197 | |||
198 | 5 | Amber Herold | 4. Leginon/Hole Targeting> While square targets are getting collected, you can |
199 | 1 | Amber Herold | start picking targets in the hole targeting menu. Pick a bunch of holes (but not |
200 | adjacent ones), then pick a focus point for rough Z- focus. To advance to the next |
||
201 | 5 | Amber Herold | square, press the play button but NOT Queue- play !http://emg.nysbc.org/software/leginon/images/icons/send_queue_out.png!. |
202 | 1 | Amber Herold | |
203 | |||
204 | 5 | Amber Herold | 5. Leginon/Hole Targeting> After the hole targets are all selected, insert the |
205 | 1 | Amber Herold | objective aperture and press Queue-play. |
206 | |||
207 | |||
208 | 5 | Amber Herold | 6. Leginon/Tomography Targeting> DON'T PRESS Queue-play until ready for final |
209 | 1 | Amber Herold | tomography acquisition. |
210 | |||
211 | |||
212 | 5 | Amber Herold | 7. Leginon/Tomography Targeting>Remember to pick a FOCUS spot for each hole and |
213 | 1 | Amber Herold | tomography target! |
214 | |||
215 | |||
216 | Tip: you can mouse over the target and get an estimate of the average intensity in |
||
217 | the popup menu, which lists X,Y,Cnts. This feature lets you rapidly screen for (1) |
||
218 | squares with the thinnest ice and (2) the thinnest targets. |
||
219 | |||
220 | |||
221 | 5 | Amber Herold | 8. Leginon/Tomography Targeting> To preview target, use preview crosshair to pick |
222 | 1 | Amber Herold | the target: |
223 | * SET defocus in preview preset to -1.2 e-06 m and adjust the exposure time (in |
||
224 | milliseconds) such that the dose is ~ 0.2 - 0.5 e-/Angstrum^2 - VERY |
||
225 | IMPORTANT! |
||
226 | * If you pick a preview target and press Play button, it will be processed |
||
227 | immediately. Therefore, don't do it until all hl images are taken. |
||
228 | * After you have inspected the preview image, you can come back to Tomography |
||
229 | Targeting node to pick final targets on the same image. Alternatively, you can |
||
230 | pick the targets first with the preview target and then remove the unwanted ones |
||
231 | after previewing. |
||
232 | |||
233 | |||
234 | |||
235 | |||
236 | |||
237 | 5 | Amber Herold | 9. Leginon/Tomography> check data collection parameters: dose, tilt range, tilt |
238 | interval. See [[Running the application#Import_Notes_about_Image_Intensity_Recorded_through_Tomography_Node|notes on image intensity recorded through |
||
239 | tomography]]. (If you want -60 to 60 tilts, you should ender -61 to 61 since the |
||
240 | 1 | Amber Herold | limiting algorithm is "smaller than" not "smaller than or equal to". |
241 | |||
242 | |||
243 | 5 | Amber Herold | 10. Leginon/Presets> setup data collection parameters for focus & tomography: |
244 | 1 | Amber Herold | mag, defocus, INT (INT for focus should be greater than for tomo). |
245 | |||
246 | |||
247 | |||
248 | |||
249 | |||
250 | |||
251 | |||
252 | h2. Final checklist: Getting ready to collect |
||
253 | |||
254 | |||
255 | |||
256 | * Leginon/Presets manager> tomo preset to scope. |
||
257 | |||
258 | |||
259 | * Tecnai/TUI/Stage> Go to reference point. |
||
260 | |||
261 | |||
262 | * Leginon/Tomography node> re-check dose and re-center beam. |
||
263 | |||
264 | |||
265 | * Leginon/Exposure targeting> press queue-play. Atlas does NOT need to be |
||
266 | reloaded. |
||
267 | |||
268 | |||
269 | |||
270 | |||
271 | |||
272 | |||
273 | |||
274 | |||
275 | |||
276 | ______ |
||
277 | |||
278 | [[Running the application|< Running the application]] | [[Leginon "MSI-Tomography" Application|Leginon "MSI-Tomography" Application ^]] |
||
279 | |||
280 | ______ |