Project

General

Profile

Sub-3Å proteasome stage-shift data collection » History » Revision 5

Revision 4 (Anchi Cheng, 08/16/2016 04:43 PM) → Revision 5/7 (Anchi Cheng, 08/16/2016 05:19 PM)

h1. Sub-3Å proteasome stage-shift data collection 

 h2. Instruments 

 # Scope: FEI Titan-Krios 
 # Camera: Gatan K2. mount directly. 1.21 Å at 29kx scope nominal mag 

 h2. Grid Preset Design 

 * 1.2/1.3 C-flat carbon support; 400 mesh Copper grid. 

 {{include(Krios-K2-TSRI)}} 

 h2. Application: 

 MSI-T2 Advanced (= MSI-T + "Beam Tilt Image" for coma-free alignment + "N2 Filling" for Krios auto nitrogen filler) 

 h2. Presets Manager: 

 No preset cycling (Krios has its own normalization). 

 h2. Queuing usage: 

 Hole Targeting: Queuing, target at the carbon area to cover four holes. 
 Exposure Targeting: Queuing. manual targeting. 

 h2. Move method to reach the targets: 

 Hole: Presets Manager stage position movement 
 Focus:    Presets Manager stage position movement 
 Exposure: [[Iterative_Stage_Movement|Navigator Iterative movement]]: Target Tolerance: 1e-7 m; Acceptable Tolerance: 3e-7 m 

 h2. Wait time before final exposure: 

 40 seconds