Project

General

Profile

Sub-3Å proteasome stage-shift data collection » History » Version 7

Anchi Cheng, 08/16/2016 05:25 PM

1 1 Anchi Cheng
h1. Sub-3Å proteasome stage-shift data collection
2
3
h2. Instruments
4
5 7 Anchi Cheng
TSRI Krios
6
# Scope: FEI Titan-Krios, 300 kV, micro probe
7 1 Anchi Cheng
# Camera: Gatan K2. mount directly. 1.21 Å at 29kx scope nominal mag
8
9 5 Anchi Cheng
h2. Grid
10 1 Anchi Cheng
11 5 Anchi Cheng
* 1.2/1.3 C-flat carbon support; 400 mesh Copper grid.
12
13 1 Anchi Cheng
{{include(Krios-K2-TSRI)}}
14 5 Anchi Cheng
15
h2. Application:
16
17
MSI-T2 Advanced (= MSI-T + "Beam Tilt Image" for coma-free alignment + "N2 Filling" for Krios auto nitrogen filler)
18
19
h2. Presets Manager:
20
21
No preset cycling (Krios has its own normalization).
22
23
h2. Queuing usage:
24
25
Hole Targeting: Queuing, target at the carbon area to cover four holes.
26
Exposure Targeting: Queuing. manual targeting.
27
28
h2. Move method to reach the targets:
29
30 6 Anchi Cheng
Hole: Presets Manager (modeled) stage position movement
31
Focus:  Presets Manager (modeled) stage position movement
32 5 Anchi Cheng
Exposure: [[Iterative_Stage_Movement|Navigator Iterative movement]]: Target Tolerance: 1e-7 m; Acceptable Tolerance: 3e-7 m
33
34
h2. Wait time before final exposure:
35
36
40 seconds