Project

General

Profile

Actions

Bug #1569

closed

defocus of the second of two autofocus sequence with the same preset and defocus adjustment is not set

Added by Anchi Cheng about 12 years ago. Updated almost 12 years ago.

Status:
Closed
Priority:
Normal
Assignee:
Category:
Leginon subsystem
Start date:
02/29/2012
Due date:
% Done:

0%

Estimated time:
Affected Version:
Appion/Leginon 2.1.0
Show in known bugs:
No
Workaround:

Description

Since the preset is not sent again to save time, the final defocus of the first (O defocus) is used in the second sequence. This makes the second one often fails and therefore set to eucentric focus and the first effort is not taking effect.


Related issues 1 (0 open1 closed)

Has duplicate Leginon - Bug #562: Focus sequence problemClosedJim Pulokas05/25/2010

Actions
Actions #1

Updated by Anchi Cheng about 12 years ago

  • Status changed from Assigned to In Code Review
  • Assignee changed from Anchi Cheng to Jim Pulokas
Actions #2

Updated by Anchi Cheng about 12 years ago

merge the change to branch 2.1 as well since it is a critical change r16647

Actions #3

Updated by Jim Pulokas almost 12 years ago

  • Status changed from In Code Review to Closed

works.

Actions

Also available in: Atom PDF