Bug #26

closed

move by image shift should base on the matrix of the preset acquiring the data not the parent

Added by Anchi Cheng almost 15 years ago. Updated about 7 years ago.

Status:
Closed
Priority:
Normal
Assignee:
Category:
-
Start date:
01/13/2010
Due date:
% Done:

70%

Estimated time:
Affected Version:
Appion/Leginon 2.0.1
Show in known bugs:
Workaround:

Description

EMTARGET that knows only the parent preset (the preset used to create the image where the target is selected on) is now responsible for determining the image shift used to acquire data. The problem is that the image shift coil values are not equivalent between different modes of microscope like between LM, M, SA. HM. There are small differences even within the same mode. This need to be changed to improve targeting accuracy.


Related issues 1 (0 open1 closed)

Related to Leginon - Feature #3909: account for image rotation in pixel to pixel transformation between magsClosedAnchi Cheng01/26/2016