Feature #10612
closedUse precalibrated ctf estimate for tomography defocus setting
Start date:
07/26/2021
Due date:
% Done:
0%
Estimated time:
Deliverable:
Description
Today's goniometer should be reproducible enough.
Updated by Anchi Cheng about 3 years ago
1. Collect tilt series using data collection settings on specimen that gives strong Thon rings. Exposure time can be adjusted to give good image.
2. If this is a movie-capable camera, align the frames to get better images.
3. estimate CTF using appion
4. Download CTF data of the the tilt series using myamiweb
5. Run leginon/tomography/upload_ctf.py to upload the ctf value difference as calibration.
6. If there are calibration at the given tem instrument, it will be used as extra correction on top of the UCSF model.
Updated by Anchi Cheng about 2 years ago
- Status changed from Assigned to Closed
Tested with success, but not widely used.