Bug #2342

closed

Dose measurement does not work correctly on binned image on Tietz F416

Added by Anchi Cheng over 11 years ago. Updated almost 7 years ago.

Status:
Closed
Priority:
High
Assignee:
Category:
-
Target version:
Start date:
04/26/2013
Due date:
% Done:

0%

Estimated time:
Affected Version:
Appion/Leginon 2.2.0
Show in known bugs:
No
Workaround:

Description

Like Eagle camera, F416 gives average of the intensity in the binned pixel while typical camera we worked with give sum.


Related issues 1 (0 open1 closed)

Related to Leginon - Bug #1764: pixel counts not always proportional to the square of binning valueClosedAnchi Cheng04/27/2012

Actions
Actions #1

Updated by Anchi Cheng over 11 years ago

  • Subject changed from Dose measurement does not work correctly on binned image on Tietz F416 and Eagle to Dose measurement does not work correctly on binned image on Tietz F416
  • Status changed from Assigned to In Code Review
  • Assignee changed from Anchi Cheng to Jim Pulokas
  • Affected Version changed from Appion/Leginon 2.1.0 to Appion/Leginon 2.2.0

r17540 fixed this for tietz.F416 and simccdcamera2.SimCCDCamera

I will put it on Tecnai2. You can assign Jeff to test.

Actions #2

Updated by Anchi Cheng almost 7 years ago

  • Status changed from In Code Review to Closed
  • Target version set to myami-3.0