Bug #2342
closedDose measurement does not work correctly on binned image on Tietz F416
Start date:
04/26/2013
Due date:
% Done:
0%
Estimated time:
Affected Version:
Appion/Leginon 2.2.0
Show in known bugs:
No
Workaround:
Description
Like Eagle camera, F416 gives average of the intensity in the binned pixel while typical camera we worked with give sum.
Updated by Anchi Cheng over 11 years ago
- Subject changed from Dose measurement does not work correctly on binned image on Tietz F416 and Eagle to Dose measurement does not work correctly on binned image on Tietz F416
- Status changed from Assigned to In Code Review
- Assignee changed from Anchi Cheng to Jim Pulokas
- Affected Version changed from Appion/Leginon 2.1.0 to Appion/Leginon 2.2.0
r17540 fixed this for tietz.F416 and simccdcamera2.SimCCDCamera
I will put it on Tecnai2. You can assign Jeff to test.
Updated by Anchi Cheng almost 7 years ago
- Status changed from In Code Review to Closed
- Target version set to myami-3.0