Project

General

Profile

Actions

Bug #2683

closed

measurement of stage z from Eucentric height is wrong when tilt backlash is significant

Added by Anchi Cheng almost 11 years ago. Updated over 10 years ago.

Status:
Closed
Priority:
Normal
Assignee:
Category:
-
Target version:
-
Start date:
03/07/2014
Due date:
% Done:

0%

Estimated time:
Affected Version:
Appion/Leginon 2.2.0
Show in known bugs:
No
Workaround:

Description

The smallest distance between + and - tilt does not correspond to eucentric point. This is most obvious on Krios


Related issues 1 (0 open1 closed)

Related to Leginon - Bug #2712: eucentric point determination by stage tilt does not work well when there is big jump in tilt direction changeClosedAnchi Cheng04/10/2014

Actions
Actions #1

Updated by Anchi Cheng almost 11 years ago

r18142 fixes this might be worth while merging to 3.0

Actions #2

Updated by Anchi Cheng over 10 years ago

  • Related to Bug #2712: eucentric point determination by stage tilt does not work well when there is big jump in tilt direction change added
Actions #3

Updated by Anchi Cheng over 10 years ago

  • Status changed from New to Closed
  • Assignee set to Anchi Cheng
  • Affected Version changed from Appion/Leginon 2.1.0 to Appion/Leginon 2.2.0

Has been used for a while, consider fixed.