Actions
Bug #2987
closedjeolcom.py does not set eucentric focus properly
Start date:
12/04/2014
Due date:
% Done:
0%
Estimated time:
Affected Version:
Appion/Leginon 3.2
Show in known bugs:
No
Workaround:
Description
Found at Wadsworth Center JEM3200/FSC/PP with K2 camera
With this combination magnifications need to go below 250x for sq mag.
In such a range zeroOM changes with mag
Updated by Anchi Cheng about 10 years ago
- Subject changed from jeolcom.py does set eucentric focus properly to jeolcom.py does not set eucentric focus properly
Updated by Anchi Cheng about 9 years ago
- Status changed from New to Closed
- Target version set to Appion/Leginon 3.2
- Affected Version changed from Appion/Leginon 3.1.0 to Appion/Leginon 3.2
Actions