Project

General

Profile

Actions

Feature #338

closed

automated coma-free beam tilt alignment during experiment

Added by Anchi Cheng over 14 years ago. Updated almost 7 years ago.

Status:
Closed
Priority:
Normal
Assignee:
Category:
-
Start date:
04/08/2010
Due date:
% Done:

0%

Estimated time:
Deliverable:

Description

Need beam tilt measured, corrected, and saved periodically during an experiment. A start for Alignment Manager.

Actions #1

Updated by Anchi Cheng over 14 years ago

First part: Jim made Alignment Manager from Target Repeater. Revision r13443, r13453, r13454

Actions #2

Updated by Anchi Cheng over 14 years ago

BeamTiltFixer added as subclass of Acquisition. BeamTiltImager has too many options for this purpose. refs Revision r13471

Actions #3

Updated by Amber Herold over 14 years ago

  • Target version changed from 8 to Appion/Leginon 2.0.0
Actions #4

Updated by Sargis Dallakyan almost 7 years ago

  • Status changed from Assigned to Closed
Actions #5

Updated by Anchi Cheng almost 7 years ago

This all works but the stigmatism estimation is not reliable enough ( need to be accurate > 95 % of time) for it to be worth doing it, even just once. Close for now.