Falcon3 implementation

Added by Anchi Cheng over 7 years ago. Updated over 5 years ago.

Status:
Closed
Priority:
Normal
Assignee:
Category:
-
Start date:
04/28/2017
Due date:
% Done:

0%

Estimated time:
Deliverable:

Description

Falcon3 has a few differences from Falcon2.

1. Acquiring cropped image is no longer allowed.
2. There is no frame configuration file for us to control frame saving.

Actions #1

Updated by Anchi Cheng almost 7 years ago

  • Target version set to Appion/Leginon 3.4

Advanced TEM Scripting is released with TItan software 2.9.1 even though the activation is still problematic.

d853d4380 has a first version of the implementation. A scripting simulator is also added for offline testing.

Actions #2

Updated by Anchi Cheng over 6 years ago

Current state. First data collection tested. Workflow works but noticed ocasional failed communication. Need more testing to debug.

Actions #3

Updated by Anchi Cheng over 5 years ago

  • Status changed from Assigned to Closed

Tested in 1.12 and 1.13 of Talos column software.

Falcon3 image server may have a timeout error which means the image does not come back. The system does recover from the error by itself. We allow it to retry in 5714813a6