Project

General

Profile

Actions

Feature #5367

closed

Option to force normalize all lens after magnification change

Added by Anchi Cheng almost 7 years ago. Updated almost 7 years ago.

Status:
Closed
Priority:
Normal
Assignee:
Category:
-
Start date:
12/18/2017
Due date:
% Done:

0%

Estimated time:
Deliverable:

Description

This feature was added for a user when his Artica does not apply normalization assigned in TUI when Leginon change magnifications in TEM Scripting, something we have not seen. It may be a change in TEM software.

To check if your installation need this, send presets at different magnification with normalization activated in TUI for magnification change and with the main screen down to observe if normalization is indeed performed.

Actions #1

Updated by Anchi Cheng almost 7 years ago

Usage:

To activate this, change fei.cfg (Most likely in C:\Program Files (x86)\myami\)

FORCE_NORMALIZE_ALL_LENS_AFTER_MAG_SETTING = False

To
FORCE_NORMALIZE_ALL_LENS_AFTER_MAG_SETTING = True

Actions #2

Updated by Anchi Cheng almost 7 years ago

  • Status changed from Assigned to Closed
Actions

Also available in: Atom PDF