Actions
Feature #5367
closedOption to force normalize all lens after magnification change
Status:
Closed
Priority:
Normal
Assignee:
Category:
-
Target version:
Start date:
12/18/2017
Due date:
% Done:
0%
Estimated time:
Deliverable:
Description
This feature was added for a user when his Artica does not apply normalization assigned in TUI when Leginon change magnifications in TEM Scripting, something we have not seen. It may be a change in TEM software.
To check if your installation need this, send presets at different magnification with normalization activated in TUI for magnification change and with the main screen down to observe if normalization is indeed performed.
Updated by Anchi Cheng almost 7 years ago
Usage:
To activate this, change fei.cfg (Most likely in C:\Program Files (x86)\myami\)
FORCE_NORMALIZE_ALL_LENS_AFTER_MAG_SETTING = False
To
FORCE_NORMALIZE_ALL_LENS_AFTER_MAG_SETTING = True
Updated by Anchi Cheng almost 7 years ago
- Status changed from Assigned to Closed