Project

General

Profile

Actions

Bug #5762

closed

AutoN2Filler node can not distinguish between slow fill and no fill COMError

Added by Anchi Cheng over 6 years ago. Updated over 6 years ago.

Status:
Closed
Priority:
Normal
Assignee:
Category:
-
Start date:
05/01/2018
Due date:
% Done:

0%

Estimated time:
Affected Version:
Appion/Leginon 3.3
Show in known bugs:
No
Workaround:

Description

FEI autofiller raises comerror in both cases after 10 minute trying. The latter is fatal but not the former. Need to check the absolute and relative level of the nitrogen to know if it is o.k. to continue collecting data.

Actions #1

Updated by Anchi Cheng over 6 years ago

Not to pause the program if the fill has reached above low trip level even if com error is returned.

Actions #2

Updated by Anchi Cheng over 6 years ago

  • Status changed from Assigned to In Test
Actions #3

Updated by Anchi Cheng over 6 years ago

  • Status changed from In Test to Closed

Works.

Actions

Also available in: Atom PDF