Bug #6338
closedThe first image shift target gets different image rotation matrix from the rest when a probe change is involved between the low mag parent preset and the acquiring preset
Status:
Closed
Priority:
Normal
Assignee:
Category:
-
Target version:
Start date:
11/19/2018
Due date:
% Done:
0%
Estimated time:
Affected Version:
Appion/Leginon 3.3
Show in known bugs:
Yes
Workaround:
Description
probe mode is not passed in the research function but obtained from the current instrument state which makes it different between the two cases
Updated by Anchi Cheng about 6 years ago
Implement what I think is a workaround by making these calibration not probe dependent. Probably better based on the preset since eucentric focuii of the two modes do not have the same objective lens current.
Updated by Anchi Cheng over 5 years ago
- Status changed from New to Closed
- Show in known bugs changed from No to Yes
This can still be a problem if some user uses nano probe and some uses micro probe at the same mag even though unlikely.