Bug #6338

closed

The first image shift target gets different image rotation matrix from the rest when a probe change is involved between the low mag parent preset and the acquiring preset

Added by Anchi Cheng about 6 years ago. Updated over 5 years ago.

Status:
Closed
Priority:
Normal
Assignee:
Category:
-
Start date:
11/19/2018
Due date:
% Done:

0%

Estimated time:
Affected Version:
Appion/Leginon 3.3
Show in known bugs:
Yes
Workaround:

Description

probe mode is not passed in the research function but obtained from the current instrument state which makes it different between the two cases

Actions #1

Updated by Anchi Cheng about 6 years ago

Implement what I think is a workaround by making these calibration not probe dependent. Probably better based on the preset since eucentric focuii of the two modes do not have the same objective lens current.

Actions #2

Updated by Anchi Cheng over 5 years ago

  • Status changed from New to Closed
  • Show in known bugs changed from No to Yes

This can still be a problem if some user uses nano probe and some uses micro probe at the same mag even though unlikely.