Actions
Bug #8193
closedSome dark references takes longer than trip value for recent dark criteria
Start date:
10/23/2019
Due date:
% Done:
0%
Estimated time:
Affected Version:
Appion/Leginon 3.4
Show in known bugs:
No
Workaround:
Description
Users report an example of TVIPS F416 with 10 repeats of dark images and on both channels. Even though individuals are within the trip level, it needs double the time to pass the trip level with two channels.
Updated by Anchi Cheng about 5 years ago
After examining the database and the leginon computer running the session, I found that the clock on the latter is about 1 minute faster. With the current trip level of 100 second, it is easily timed out.
Updated by Anchi Cheng almost 5 years ago
- Status changed from Assigned to Closed
No problem since the fix.
Actions