Acquiring summed frame image thru Leginon » History » Version 4
Anchi Cheng, 06/18/2012 08:43 PM
1 | 1 | Anchi Cheng | h1. Acquiring summed frame image thru Leginon |
---|---|---|---|
2 | |||
3 | h2. Camera Configuration |
||
4 | 2 | Anchi Cheng | |
5 | 1 | Anchi Cheng | Refer to [[DE-12_introduction#Camera_Settings_for_DE-12|introduction]] on the options |
6 | # Select DE12 as the camera |
||
7 | # Uncheck "save raw frames" |
||
8 | # Leave "Frames to use" empty to sum up all frames |
||
9 | |||
10 | h2. acquire [[Bright and Dark reference images]] |
||
11 | |||
12 | ** Leginon does not research references by frame rate and sensor temperature. If you change either of them, you will need to acquire new references. |
||
13 | ** Use valid exposure time (multiple of frame time length) |
||
14 | ** Use long positive readout delay time (>= 2 frame length) to ensure uniform intensity readout on the first recorded frame. |
||
15 | ** It does not matter whether you save raw frames or not in this process. Leginon does not use them for corrections. |
||
16 | |||
17 | 4 | Anchi Cheng | h3. typical number of dose rate, number of frames, number of images used in reference image acquisition at NRAMM |
18 | |||
19 | We have been using the following numbers. However, we did not do a systematic study on what is required, so it might not be optimal: |
||
20 | * dose rate: 4 electron per pixel per frame (similar to our typical dose rate for experiment) |
||
21 | * number of frames in summed image: 10 (i.e. 407 ms if using the default 24.57 frames/second) |
||
22 | * number of images for averaging: 20 |
||
23 | |||
24 | 3 | Anchi Cheng | h2. Usage |
25 | |||
26 | # Select DE12 as the camera, configure and uncheck "save raw frames" in the camera configuration of the preset that you want to use and go for it. |