Project

General

Profile

Adding DE-12 to Leginon » History » Revision 11

Revision 10 (Anchi Cheng, 01/09/2014 09:32 PM) → Revision 11/16 (Anchi Cheng, 01/09/2014 09:33 PM)

h1. Installation Guide for DE-12 or DE-20 

 h2.    DE-12 is on a computer separated from the microscope 

 Please read [[Using_Leginon_on_a_system_where_the_microscope_and_camera_are_controlled_by_different_computers]] first. 

 h2. Extra Package and Installation 

 None. 

 h2. instruments.cfg 

 {{include(Instrument.cfg for DE-12)}} 
 height and width should be set according to the model. 

 h2. Setup 

 * Create the folder to store the dose fractionation raw frames.    Default location is D:\\frames. 
 * Setup [[DDD_raw_frame_file_transfer|raw frame file transfer]] from a network data server. 

 h2. Testing 

 h2. Testing with pyscope 

 {{include(DE-12 Installation Pyscope Test)}} 

 h2. Programs to open before Leginon Client: DE Server 

 See how to use DE camera usage in 

 
  * [[DE-12 introduction]] Please read this!!! 
 
  * [[Acquiring summed frame image thru Leginon]] 
 
  * [[DDD raw frame saving|Triggering raw frame saving]] 
 
  * [[DDD raw frame file transfer|Transferring raw frame images]] off the solid-state drive of DE-12 during data collection to a network drive. 
 
  * [[Make DE-12 movie stack|Compiling corrected movie stack]] with Appion using bright and dark images acquired in Leginon