Adding DE-12 to Leginon » History » Revision 13
Revision 12 (Anchi Cheng, 04/11/2014 08:48 PM) → Revision 13/16 (Anchi Cheng, 02/02/2016 02:30 PM)
h1. Installation Guide for DE-12 or DE-20
h2. DE-12 is on a computer separated from the microscope
Please read [[Using_Leginon_on_a_system_where_the_microscope_and_camera_are_controlled_by_different_computers]] first.
h2. Extra Package and Installation
|_.program/package|_.notes|
|DirectElectronAPI.exe|(Should come with the camera)|
|"Google protobuf":http://code.google.com/p/protobuf/||
{{include(portobuf installation instruction)}}
h2. instruments.cfg
{{include(Instrument.cfg for DE-12)}}
height and width should be set according to the model.
h2. Setup
* Create the folder to store the dose fractionation raw frames. Default location is D:\\frames.
* Setup [[DDD_raw_frame_file_transfer|raw frame file transfer]] from a network data server.
h2. Testing
h2. Testing with pyscope
{{include(DE-12 Installation Pyscope Test)}}
h2. Programs to open before Leginon Client: DE Server
See how to use DE camera usage in
* [[DE-12 introduction]] Please read this!!!
* [[Acquiring summed frame image thru Leginon]]
* [[DDD raw frame saving|Triggering raw frame saving]]
* [[DDD raw frame file transfer|Transferring raw frame images]] off the solid-state drive of DE-12 during data collection to a network drive.
* [[Make DE-12 movie stack|Compiling corrected movie stack]] with Appion using bright and dark images acquired in Leginon