AlignZeroLoss » History » Version 1

Anchi Cheng, 12/06/2018 10:19 PM

1 1 Anchi Cheng
h1. Align Zero Loss Node Class
2
3
AlignZeroLoss class is used to align the energy filter slit so that zero loss peak is allowed to pass through optimally. It is used with Gatan Energy Filter such as Bioquantum series.  At given interval and optionally below an intensity threshold, the alignment process is initiated.
4
5
h2. Reference target requirement:
6
7
The alignment requires an area that either empty or with little obstruction.  This reference target stage position must be defined for the node to function.
8
9
h2. Bindings
10
11
*Required Binding for initiate the time interval check and alignment*
12
13
[[Acquisition]] -> AlignZeroLossPeakPublishEvent -> Align Zero Loss Peak
14
15
*When [[Phase_Plate_Aligner_Node_Class|phase plate aligner]] is used for phase plate, bind phase plate aligner with align zeo loss peak to force zero loss peak alignment at each phase plate patch change.
16
17
[[Phase_Plate_Aligner_Node_Class|Phase Plate Aligner]] -> FixAlignment -> Align Zero Loss Peak
18
 
19
*Required Binding for moving to reference target*
20
21
Align Zero Loss Peak ->ChangePresetEvent -> Presets Manager
22
Presets Manger -> PresetChangedEvent -> Align Zero Loss Peak
23
Align Zero Loss Peak -> MoveToTargetEvent -> Navigator
24
Navigator -> MoveToTargetDoneEvent -> Align Zero Loss Peak
25
26
h2. Reference Target Settings
27
28
* Bypass Conditioner should be deactivated for the node to be used.
29
30
h2. Check Zero Loss Peak Shift Settings
31
32
* Set the standard deviation threshold to 0 if you want it to do alignment based only on time interval.