Project

General

Profile

Actions

Aperture selection through tem controller » History » Revision 2

« Previous | Revision 2/5 (diff) | Next »
Anchi Cheng, 04/24/2020 03:39 PM


Aperture selection through tem controller

This feature allows selection of aperture of TFS microscopes with automated apertures. The current implementation uses AutoIt Scripting and reads the configuration from c:\Program Files (x86)\myami\fei.cfg

Setup

compile autoit script

The script is pyscope/autoit/ApertureSelection.au3
Use AutoIt suite to compile it. Either 32 or 64-bit version works as far as I know. Save the executable at the place you plan to specify in fei.cfg

Setup pyscope fei.cfg

Copy from pyscope/fei.cfg.template these lines to fei.cfg used on your scope. In standard installation, this should be "c:\Program Files (x86)\myami\fei.cfg"

Modify these settings according to your scope.

[aperture]
# Disable control of auto apertures if not working or not available
# not all versions of microscope software allows auto aperture control
# from scripting level.
USE_AUTO_APERTURE = False
# AutoIt Automation
AUTOIT_APERTURE_SELECTION_EXE_PATH = c:\Users\supervisor\Desktop\Anchi\AutoItScripts\ApertureSelection.exe
# Does the Titan column has automated c3 aperture ?  This affects the id of the
# gui button for AutoIt script. 
HAS_AUTO_C3 = True
# Individual aperture selections.  If retractable, add open at the end of the list
# List these in the order of what appears in the TUI
# Note: can not handle multiple apertures of the same name.
CONDENSER_2 = 150, 100, 50, 20
OBJECTIVE = 100, 70, 50, open
# Optional selected area aperture control. comment or remove the line to deactivate it.
SELECTED_AREA = 200,100, 40, 10, open

Testing

Testing the AutoIt executable

  1. Open TEM user interface.
  2. Retract the objective aperture.
  3. Double click Autoit executable to run.
    If this works right, it should insert the 100 um aperture.

Testing in with python command from pyscope

from pyscope import fei
f=fei.Krios()  # Change this to match your instrument class

f.getApertureSelections('objective') # This should give you the list you specified in fei.cfg
'100', '70', '50' , 'open'

f.getApertureSelection('objective')  # The returned value should be the current aperture position as a string
'100'

f.setApertureSelection('objective', '50') # This should insert the 50 um aperture

f.retractApertureMechanism('objective') # This should retract the aperture as if you click on the Activate/Deactivate aperture button to disable it.

Updated by Anchi Cheng over 4 years ago · 2 revisions