Autofocus Calibration with Beam Tilt Node » History » Version 10

Anchi Cheng, 05/01/2017 04:50 PM

1 1 Amber Herold
h1. Autofocus Calibration with Beam Tilt Node
2
3
4
5 8 Anchi Cheng
Two functions have to be performed in Beam Tilt node to perform defocus and
6
z-height corrections used in autofocus procedures. Defocus calibration are
7
required for defocus correction. This plus a record of eucentric focus value are
8 1 Amber Herold
required for the z-height correction to eucentric height.
9
10 8 Anchi Cheng
*We have disabled Stigmator Calibration and Correction in 2.2 because it has not been useful*
11 1 Amber Herold
12 8 Anchi Cheng
13 3 Amber Herold
Beam Tilt Defocus/Stigmator Calibration and Eucentric Focus Recording Need for the Example MSI:
14 1 Amber Herold
15 3 Amber Herold
16 6 Anchi Cheng
|Preset|magnification|
17 1 Amber Herold
|hl|5000|
18
|fa|50000|
19
20
21
22
23
24
h2. Defocus Calibration
25
26
27 6 Anchi Cheng
#  Leginon/Presets Manager> send the preset you want to make the calibration "To Scope"
28 4 Amber Herold
 
29 9 Gabriel Lander
#  Leginon/Beam Tilt/Toolbar/Setting> check and make sure that "Override Preset" is NOT checked
30 4 Amber Herold
#  Leginon/Beam Tilt/Toolbar> select Defocus as calibration type
31
 
32
#  Leginon/Beam Tilt/Toolbar> open "Parameter Settings" window to set parameters as follows:
33
 
34 1 Amber Herold
at 50,000x: defocus1 = -2e-06, defocus2 = -4e-06 and beam tilt angle=0.01
35 4 Amber Herold
 
36 1 Amber Herold
at 5,000x: defocus1 = -2e-05, defocus2 = -4e-05 and beam tilt angle=0.01
37 4 Amber Herold
 
38
#  Leginon/Beam Tilt/Toolbar> click Calibrate.
39
 
40
#  Occasionally the beam may be partially blocked by the objective aperture and
41 1 Amber Herold
produce a dark field image upon beam tilts. Reducing the beam tilt angle, using a
42
larger objective aperture, and/or centering the objective aperture accurately are the
43
possible solutions.
44 4 Amber Herold
45 1 Amber Herold
46
47
48
49
h2.  Checking the Calibrations
50
51
52
53 8 Anchi Cheng
Once Defocus are calibrated against beam-tilt induced image shift,
54 3 Amber Herold
the two can be measured using the Measure Function in the Toolbar.
55 1 Amber Herold
56 3 Amber Herold
57 1 Amber Herold
58
59 4 Amber Herold
#  Leginon/Beam Tilt/Toolbar> click Measure to open measurement submenu.
60
 
61 8 Anchi Cheng
#  Leginon/Beam Tilt/Measure> click Measure (to test the beam tilt defocus calibration).
62 4 Amber Herold
 
63
#  Leginon/Beam Tilt/Measure> click Correct Defocus to make the correction according to the measurement.
64
 
65
#  Leginon/Beam Tilt/Measure> click Reset Defocus and exit the sub-window.
66
 
67
#  Leginon/Presets Managaer> send the fc preset "To Scope"
68
 
69
#  Leginon/Focus/Settings> select fc as the preset used for the acquisition.
70
 
71
#  Leginon/Focus/Toolbar> start Manual Focusing.
72
 
73 1 Amber Herold
Power Spectrum of the image at defocus specified by "fc" preset should appear on
74 3 Amber Herold
the screen.
75 4 Amber Herold
 
76
#  Leginon/Focus/Manual Focus> set 0 defocus to scope by selecting the movetype
77 1 Amber Herold
to "defocus" and the number entry in meter to "0" and then click "Set
78
Instrument".
79 4 Amber Herold
 
80
The Thone ring should disappear as the defocus is changed to zero measured by Beam
81
Tilt node. If not, recalibration of defocus may be needed.
82
 
83
#  Leginon/Focus/ToolbarManual Focus> stop Manual Focusing by clicking the "stop" button.
84
 
85 1 Amber Herold
#  Leginon/Beam Tilt/Toolbar> click Measure to open measurement sub-menu.
86 4 Amber Herold
 
87
#  Leginon/Focus/Toolbar> start Manual Focusing and check to see if the
88 1 Amber Herold
correction makes bad stigmation better.
89
If not, redo stigmation correction.
90
91
92 3 Amber Herold
93 1 Amber Herold
Trouble Shooting:
94
95 10 Anchi Cheng
See [[Optimizing_Autofocus]] for some pictures and more detailed description.
96 1 Amber Herold
97
*  The microscope must be well aligned. The most critical are beam tilt pivot point and rotation center, both should be aligned at eucentric focus. Please refer to [[Microscope Set-up|Microscope Set-up]] to align the microscope well.
98
99
100
*  Chose a beam tilt and defocus change that will give a 10-20% shift in the image.
101 3 Amber Herold
102 1 Amber Herold
103
*  The beam should cover the entire CCD imaging area.
104 3 Amber Herold
105 1 Amber Herold
106 3 Amber Herold
*  Have an easily recognized object at the center of the image for easy correlation. Avoid uneven areas.
107 1 Amber Herold
108
109
*  Watch the images movement. There should not be any contrast reversal or appearance of the edge of the beam in any image acquired. Contrast reversal is caused by misaligned objective aperture or by too large a beam tilt angle so that part of the beam is blocked by the aperture when tilted.
110
111
112
*  Check the correlation image and peak location to find out if the image contrast is too low for good correlations.
113
114
h2.  Eucentric Focus Storage and Retrieval
115
116 3 Amber Herold
Encentric Focus From/To Scope (From/To Scope icons) do what they indicate. The "from
117 1 Amber Herold
Scope" function simply records the current absolute focus value at the scope.
118
Therefore,
119 3 Amber Herold
120 4 Amber Herold
#  Scope> eucenter the grid and reset defocus at U-centric height at the required high tension and at a high mag where focusing is easier.
121
 
122
#  Leginon/Presets Manager> select the preset that will perform z-focusing (i.e., hl). temporarily change its defocus to 0.0
123
 
124
#  Leginon/Presets Manager> sent the preset "To Scope"
125
 
126 5 Anchi Cheng
#  Leginon/Beam Tilt> left-click the Encentric Focus From Scope icon to save the focus value.
127 4 Amber Herold
 
128
#  Leginon/Presets Manager> return the defocus value back to its original.
129 1 Amber Herold
130
h2.  Rotation Center Storage and Retrieval
131 3 Amber Herold
132 1 Amber Herold
Rotation Center From/To Scope (From/To Scope icons) do what they indicate. The "from Scope" function simply records the current beam tilt value at the scope. Therefore,
133
134 3 Amber Herold
135 4 Amber Herold
#  Scope> Align the rotation center at a magnification and high tension where the value will be used.
136
 
137 5 Anchi Cheng
#  Leginon/Beam Tilt> left-click the Rotation Center From Scope icon to save the beam tilt value to Leginon database.
138 2 Amber Herold
139
140
______
141
142
[[Checking Matrix and Modeled Stage Position Calibration|< Checking Matrix and Modeled Stage Position Calibration]] | [[Dose (Camera Sensitivity) Calibration|Dose (Camera Sensitivity) Calibration >]]
143
144
______