Autofocus Calibration with Beam Tilt Node » History » Version 2

Amber Herold, 04/27/2010 10:50 AM

1 1 Amber Herold
h1. Autofocus Calibration with Beam Tilt Node
2
3
4
5
Three functions have to be performed in Beam Tilt node to perform defocus, stigmation and
6
z-height corrections used in autofocus procedures. Defocus and stigmator calibration are
7
required for defocus/stigmation correction. These plus a record of eucentric focus value are
8
required for the z-height correction to eucentric height.
9
10
11
12
|<filename>Preset</filename>|<filename>magnification</filename>|
13
|hl|5000|
14
|fa|50000|
15
16
17
18
19
20
h2. Defocus Calibration
21
22
23
24
25
26
*  Leginon/Presets Managaer> send the preset you want to make the calibration "To
27
Scope"
28
29
30
*  Leginon/Beam Tilt/Toolbar/Setting> check and make sure that it does NOT
31
overwrite presets
32
33
34
*  Leginon/Beam Tilt/Toolbar> select Defocus as calibration type
35
36
37
*  Leginon/Beam Tilt/Toolbar> open "Parameter Settings" window to set parameters
38
as follows:
39
40
41
at 50,000x: defocus1 = -2e-06, defocus2 = -4e-06 and beam tilt angle=0.01
42
43
44
at 5,000x: defocus1 = -2e-05, defocus2 = -4e-05 and beam tilt angle=0.01
45
46
47
*  Leginon/Beam Tilt/Toolbar> click Calibrate.
48
49
50
*  Occasionally the beam may be partially blocked by the objective aperture and
51
produce a dark field image upon beam tilts. Reducing the beam tilt angle, using a
52
larger objective aperture, and/or centering the objective aperture accurately are the
53
possible solutions.
54
55
56
57
58
59
60
61
62
63
h2.  Beam Tilt Stigmator calibration
64
65
66
67
The Beam Tilt Stigmator calibration is used by Leginon to correct astigmatism. This
68
calibration functions well if the microscope has been well aligned first. Stigmation
69
correction at low mag such as 5000x is not necessary practically but the calibration is
70
still required for the equation used in autofocusing.
71
72
73
74
75
*  Leginon/Presets Managaer> if hasn't done so, send the preset you want to make
76
the calibration "To Scope"
77
78
79
*  Leginon/Beam Tilt/Toolbar> select Stigmators for calibration.
80
81
82
*  Leginon/Beam Tilt/Toolbar> open "Parameter Settings" window to set parameters
83
as follows:
84
85
86
at 50,000x: beam tilt angle= 0.01 and delta stig.= 0.2
87
88
89
at 5,000x: beam tilt angle= 0.01 and delta stig.= 0.2
90
91
92
*  Leginon/Beam Tilt/Toolbar> click Calibrate.
93
94
95
*  Occasionally the beam may be partially blocked by the objective aperture and
96
produce a dark field image upon beam tilts. Reducing the beam tilt angle, using a
97
larger objective aperture, and/or centering the objective aperture accurately are the
98
possible solutions.
99
100
101
102
103
104
105
106
107
108
h2.  Checking the Calibrations
109
110
111
112
Once both Defocus and Stigmation are calibrated against beam-tilt induced image shift,
113
the two can be measured using the Measure Function in the Toolbar.
114
115
116
117
118
*  Leginon/Beam Tilt/Toolbar> click Measure to open measurement submenu.
119
120
121
*  Leginon/Beam Tilt/Measure> click Measure (to test the beam tilt defocus and
122
stigmation calibration).
123
124
125
*  Leginon/Beam Tilt/Measure> click Correct Defocus to make the correction
126
according to the measurement.
127
128
129
*  Leginon/Beam Tilt/Measure> click Reset Defocus and exit the subwindow.
130
131
132
*  Leginon/Presets Managaer> send the fc preset "To Scope"
133
134
135
*  Leginon/Focus/Settings> select fc as the preset used for the
136
acquisition.
137
138
139
*  Leginon/Focus/Toolbar> start Manual Focusing.
140
141
142
Power Spectrum of the image at defocus specified by "fc" preset should appear on
143
the screen.
144
145
146
*  Leginon/Focus/Manual Focus> set 0 defocus to scope by selecting the movetype
147
to "defocus" and the number entry in meter to "0" and then click "Set
148
Instrument".
149
150
151
The Thone ring should disappear as the defocus is changed to zero measured by Beam
152
Tilt node. If not, recalibration of defocus may be needed.
153
154
155
*  Leginon/Focus/ToolbarManual Focus> stop Manual Focusing by clicking the "stop"
156
button.
157
158
159
*  Leginon/Beam Tilt/Toolbar> click Measure to open measurement submenu.
160
161
162
*  Leginon/Beam Tilt/Measure> click Correct Stigamator to make the correction
163
according to the measurement.
164
165
166
*  Leginon/Focus/Toolbar> start Manual Focusing and check to see if the
167
correction makes bad stigmation better.
168
169
170
If not, redo stigmation correction.
171
172
173
174
175
176
177
Trouble Shooting:
178
179
180
181
182
*  The microscope must be well aligned. The most critical are beam tilt pivot point
183
and rotation center, both should be aligned at eucentric focus. Please refer to <link
184
linkend="microscope_setup">Microscope Set-up</link> to align the microscope
185
well.
186
187
188
*  Chose a beam tilt and defocus change that will give a 10-20% shift in the
189
image.
190
191
192
*  The beam should cover the entire CCD imaging area.
193
194
195
*  Have a easily reconized object at the center of the image for easy correlation.
196
Avoid uneven area.
197
198
199
*  Watch the images movement. There should not be any contrast reversal or apearance
200
of the edge of the beam in any image acquired. Contrast reversal is caused by
201
misaligned objective aperture or by too large a beam tilt angle so that part of the
202
beam is blocked by the aperture when tilted.
203
204
205
*  Check the correlation image and peak location to find out if the image contrast
206
is too low for good correlations.
207
208
209
*  The default values to correct calibration the stagmators are recommended. If
210
these values do not produce good stigmation calibration results and if the microscope
211
has been well aligned, then progressively and slowly change the Stigmator Tilt and
212
Stigmator Delta values until the images cross correlations result in measureable
213
shifts between steps of the calibration.
214
215
216
217
218
219
220
221
222
223
h2.  Eucentric Focus Storage and Retrieval
224
225
226
227
Encentric Focus From/To Scope (From/To Scope icons) do what they indicate. The "from
228
Scope" function simply records the current absolute focus value at the scope.
229
Therefore,
230
231
232
233
234
*  Scope> eucenter the grid and reset defocus at U-centric height at the required
235
high tension and at a high mag where focusing is easier.
236
237
238
*  Leginon/Presets Manager> select the preset that will perform z-focusing (i.e.,
239
hl). temporarily change its defocus to 0.0
240
241
242
*  Leginon/Presets Manager> sent the preset "To Scope"
243
244
245
*  Leginon/Beam Tilt> left-click the Encentric Focus To Scope icon to save the
246
focus value.
247
248
249
*  Leginon/Presets Manager> return the defocus value back to its original.
250
251
252
253
254
255
256
257
258
259
h2.  Rotation Center Storage and Retrieval
260
261
262
263
Rotation Center From/To Scope (From/To Scope icons) do what they indicate. The "from
264
Scope" function simply records the current beam tilt value at the scope. Therefore,
265
266
267
268
269
*  Scope> Align the rotation center at a magnification and high tension where the
270
value will be used.
271
272
273
*  Leginon/Beam Tilt> left-click the Rotation Center To Scope icon to save the
274
beam tilt value.
275 2 Amber Herold
276
277
______
278
279
[[Checking Matrix and Modeled Stage Position Calibration|< Checking Matrix and Modeled Stage Position Calibration]] | [[Dose (Camera Sensitivity) Calibration|Dose (Camera Sensitivity) Calibration >]]