h1. Autofocus Calibration with Beam Tilt Node 



 Three functions have to be performed in Beam Tilt node to perform defocus, stigmation and 
 z-height corrections used in autofocus procedures. Defocus and stigmator calibration are 
 required for defocus/stigmation correction. These plus a record of eucentric focus value are 
 required for the z-height correction to eucentric height. 



 |<filename>Preset</filename>|<filename>magnification</filename>| 
 |hl|5000| 
 |fa|50000| 





 h2. Defocus Calibration 





 *    Leginon/Presets Managaer> send the preset you want to make the calibration "To 
 Scope" 


 *    Leginon/Beam Tilt/Toolbar/Setting> check and make sure that it does NOT 
 overwrite presets 


 *    Leginon/Beam Tilt/Toolbar> select Defocus as calibration type 


 *    Leginon/Beam Tilt/Toolbar> open "Parameter Settings" window to set parameters 
 as follows: 


 at 50,000x: defocus1 = -2e-06, defocus2 = -4e-06 and beam tilt angle=0.01 


 at 5,000x: defocus1 = -2e-05, defocus2 = -4e-05 and beam tilt angle=0.01 


 *    Leginon/Beam Tilt/Toolbar> click Calibrate. 


 *    Occasionally the beam may be partially blocked by the objective aperture and 
 produce a dark field image upon beam tilts. Reducing the beam tilt angle, using a 
 larger objective aperture, and/or centering the objective aperture accurately are the 
 possible solutions. 









 h2.    Beam Tilt Stigmator calibration 



 The Beam Tilt Stigmator calibration is used by Leginon to correct astigmatism. This 
 calibration functions well if the microscope has been well aligned first. Stigmation 
 correction at low mag such as 5000x is not necessary practically but the calibration is 
 still required for the equation used in autofocusing. 




 *    Leginon/Presets Managaer> if hasn't done so, send the preset you want to make 
 the calibration "To Scope" 


 *    Leginon/Beam Tilt/Toolbar> select Stigmators for calibration. 


 *    Leginon/Beam Tilt/Toolbar> open "Parameter Settings" window to set parameters 
 as follows: 


 at 50,000x: beam tilt angle= 0.01 and delta stig.= 0.2 


 at 5,000x: beam tilt angle= 0.01 and delta stig.= 0.2 


 *    Leginon/Beam Tilt/Toolbar> click Calibrate. 


 *    Occasionally the beam may be partially blocked by the objective aperture and 
 produce a dark field image upon beam tilts. Reducing the beam tilt angle, using a 
 larger objective aperture, and/or centering the objective aperture accurately are the 
 possible solutions. 









 h2.    Checking the Calibrations 



 Once both Defocus and Stigmation are calibrated against beam-tilt induced image shift, 
 the two can be measured using the Measure Function in the Toolbar. 




 *    Leginon/Beam Tilt/Toolbar> click Measure to open measurement submenu. 


 *    Leginon/Beam Tilt/Measure> click Measure (to test the beam tilt defocus and 
 stigmation calibration). 


 *    Leginon/Beam Tilt/Measure> click Correct Defocus to make the correction 
 according to the measurement. 


 *    Leginon/Beam Tilt/Measure> click Reset Defocus and exit the subwindow. 


 *    Leginon/Presets Managaer> send the fc preset "To Scope" 


 *    Leginon/Focus/Settings> select fc as the preset used for the 
 acquisition. 


 *    Leginon/Focus/Toolbar> start Manual Focusing. 


 Power Spectrum of the image at defocus specified by "fc" preset should appear on 
 the screen. 


 *    Leginon/Focus/Manual Focus> set 0 defocus to scope by selecting the movetype 
 to "defocus" and the number entry in meter to "0" and then click "Set 
 Instrument". 


 The Thone ring should disappear as the defocus is changed to zero measured by Beam 
 Tilt node. If not, recalibration of defocus may be needed. 


 *    Leginon/Focus/ToolbarManual Focus> stop Manual Focusing by clicking the "stop" 
 button. 


 *    Leginon/Beam Tilt/Toolbar> click Measure to open measurement submenu. 


 *    Leginon/Beam Tilt/Measure> click Correct Stigamator to make the correction 
 according to the measurement. 


 *    Leginon/Focus/Toolbar> start Manual Focusing and check to see if the 
 correction makes bad stigmation better. 


 If not, redo stigmation correction. 






 Trouble Shooting: 




 *    The microscope must be well aligned. The most critical are beam tilt pivot point 
 and rotation center, both should be aligned at eucentric focus. Please refer to <link 
 linkend="microscope_setup">Microscope Set-up</link> to align the microscope 
 well. 


 *    Chose a beam tilt and defocus change that will give a 10-20% shift in the 
 image. 


 *    The beam should cover the entire CCD imaging area. 


 *    Have a easily reconized object at the center of the image for easy correlation. 
 Avoid uneven area. 


 *    Watch the images movement. There should not be any contrast reversal or apearance 
 of the edge of the beam in any image acquired. Contrast reversal is caused by 
 misaligned objective aperture or by too large a beam tilt angle so that part of the 
 beam is blocked by the aperture when tilted. 


 *    Check the correlation image and peak location to find out if the image contrast 
 is too low for good correlations. 


 *    The default values to correct calibration the stagmators are recommended. If 
 these values do not produce good stigmation calibration results and if the microscope 
 has been well aligned, then progressively and slowly change the Stigmator Tilt and 
 Stigmator Delta values until the images cross correlations result in measureable 
 shifts between steps of the calibration. 









 h2.    Eucentric Focus Storage and Retrieval 



 Encentric Focus From/To Scope (From/To Scope icons) do what they indicate. The "from 
 Scope" function simply records the current absolute focus value at the scope. 
 Therefore, 




 *    Scope> eucenter the grid and reset defocus at U-centric height at the required 
 high tension and at a high mag where focusing is easier. 


 *    Leginon/Presets Manager> select the preset that will perform z-focusing (i.e., 
 hl). temporarily change its defocus to 0.0 


 *    Leginon/Presets Manager> sent the preset "To Scope" 


 *    Leginon/Beam Tilt> left-click the Encentric Focus To Scope icon to save the 
 focus value. 


 *    Leginon/Presets Manager> return the defocus value back to its original. 









 h2.    Rotation Center Storage and Retrieval 



 Rotation Center From/To Scope (From/To Scope icons) do what they indicate. The "from 
 Scope" function simply records the current beam tilt value at the scope. Therefore, 




 *    Scope> Align the rotation center at a magnification and high tension where the 
 value will be used. 


 *    Leginon/Beam Tilt> left-click the Rotation Center To Scope icon to save the 
 beam tilt value. 


 ______ 

 [[Checking Matrix and Modeled Stage Position Calibration|< Checking Matrix and Modeled Stage Position Calibration]] | [[Dose (Camera Sensitivity) Calibration|Dose (Camera Sensitivity) Calibration >]] 