3 | |||
4 | |||
5 | Three functions have to be performed in Beam Tilt node to perform defocus, stigmation and |
||
6 | z-height corrections used in autofocus procedures. Defocus and stigmator calibration are |
||
7 | required for defocus/stigmation correction. These plus a record of eucentric focus value are |
||
8 | required for the z-height correction to eucentric height. |
||
9 | |||
10 | |||
11 | 3 | Amber Herold | Beam Tilt Defocus/Stigmator Calibration and Eucentric Focus Recording Need for the Example MSI: |
12 | 1 | Amber Herold | |
13 | 3 | Amber Herold | |
14 | 1 | Amber Herold | |<filename>Preset</filename>|<filename>magnification</filename>| |
15 | |hl|5000| |
||
16 | |fa|50000| |
||
17 | |||
18 | |||
19 | |||
20 | |||
21 | |||
22 | h2. Defocus Calibration |
||
23 | |||
24 | |||
25 | |||
26 | |||
27 | |||
28 | 3 | Amber Herold | * Leginon/Presets Managaer> send the preset you want to make the calibration "To Scope" |
29 | 1 | Amber Herold | |
30 | |||
31 | 3 | Amber Herold | * Leginon/Beam Tilt/Toolbar/Setting> check and make sure that it does NOT overwrite presets |
32 | 1 | Amber Herold | |
33 | |||
34 | * Leginon/Beam Tilt/Toolbar> select Defocus as calibration type |
||
35 | |||
36 | |||
37 | 3 | Amber Herold | * Leginon/Beam Tilt/Toolbar> open "Parameter Settings" window to set parameters as follows: |
38 | 1 | Amber Herold | at 50,000x: defocus1 = -2e-06, defocus2 = -4e-06 and beam tilt angle=0.01 |
39 | at 5,000x: defocus1 = -2e-05, defocus2 = -4e-05 and beam tilt angle=0.01 |
||
40 | |||
41 | |||
42 | * Leginon/Beam Tilt/Toolbar> click Calibrate. |
||
43 | |||
44 | |||
45 | * Occasionally the beam may be partially blocked by the objective aperture and |
||
46 | produce a dark field image upon beam tilts. Reducing the beam tilt angle, using a |
||
47 | larger objective aperture, and/or centering the objective aperture accurately are the |
||
48 | possible solutions. |
||
49 | |||
50 | |||
51 | |||
52 | |||
53 | h2. Beam Tilt Stigmator calibration |
||
54 | |||
55 | |||
56 | |||
57 | The Beam Tilt Stigmator calibration is used by Leginon to correct astigmatism. This |
||
58 | calibration functions well if the microscope has been well aligned first. Stigmation |
||
59 | correction at low mag such as 5000x is not necessary practically but the calibration is |
||
60 | still required for the equation used in autofocusing. |
||
61 | |||
62 | |||
63 | |||
64 | |||
65 | * Leginon/Presets Managaer> if hasn't done so, send the preset you want to make |
||
66 | the calibration "To Scope" |
||
67 | |||
68 | |||
69 | * Leginon/Beam Tilt/Toolbar> select Stigmators for calibration. |
||
70 | |||
71 | |||
72 | * Leginon/Beam Tilt/Toolbar> open "Parameter Settings" window to set parameters |
||
73 | as follows: |
||
74 | at 50,000x: beam tilt angle= 0.01 and delta stig.= 0.2 |
||
75 | at 5,000x: beam tilt angle= 0.01 and delta stig.= 0.2 |
||
76 | |||
77 | |||
78 | * Leginon/Beam Tilt/Toolbar> click Calibrate. |
||
79 | |||
80 | |||
81 | * Occasionally the beam may be partially blocked by the objective aperture and |
||
82 | produce a dark field image upon beam tilts. Reducing the beam tilt angle, using a |
||
83 | larger objective aperture, and/or centering the objective aperture accurately are the |
||
84 | possible solutions. |
||
85 | |||
86 | |||
87 | |||
88 | h2. Checking the Calibrations |
||
89 | |||
90 | |||
91 | |||
92 | Once both Defocus and Stigmation are calibrated against beam-tilt induced image shift, |
||
93 | the two can be measured using the Measure Function in the Toolbar. |
||
94 | |||
95 | |||
96 | |||
97 | |||
98 | * Leginon/Beam Tilt/Toolbar> click Measure to open measurement submenu. |
||
99 | |||
100 | |||
101 | 3 | Amber Herold | * Leginon/Beam Tilt/Measure> click Measure (to test the beam tilt defocus and stigmation calibration). |
102 | 1 | Amber Herold | |
103 | |||
104 | 3 | Amber Herold | * Leginon/Beam Tilt/Measure> click Correct Defocus to make the correction according to the measurement. |
105 | 1 | Amber Herold | |
106 | |||
107 | 3 | Amber Herold | * Leginon/Beam Tilt/Measure> click Reset Defocus and exit the sub-window. |
108 | 1 | Amber Herold | |
109 | |||
110 | * Leginon/Presets Managaer> send the fc preset "To Scope" |
||
111 | |||
112 | |||
113 | 3 | Amber Herold | * Leginon/Focus/Settings> select fc as the preset used for the acquisition. |
114 | 1 | Amber Herold | |
115 | |||
116 | * Leginon/Focus/Toolbar> start Manual Focusing. |
||
117 | Power Spectrum of the image at defocus specified by "fc" preset should appear on |
||
118 | the screen. |
||
119 | |||
120 | |||
121 | * Leginon/Focus/Manual Focus> set 0 defocus to scope by selecting the movetype |
||
122 | to "defocus" and the number entry in meter to "0" and then click "Set |
||
123 | Instrument". |
||
124 | The Thone ring should disappear as the defocus is changed to zero measured by Beam |
||
125 | Tilt node. If not, recalibration of defocus may be needed. |
||
126 | |||
127 | |||
128 | 3 | Amber Herold | * Leginon/Focus/ToolbarManual Focus> stop Manual Focusing by clicking the "stop" button. |
129 | 1 | Amber Herold | |
130 | |||
131 | 3 | Amber Herold | * Leginon/Beam Tilt/Toolbar> click Measure to open measurement sub-menu. |
132 | 1 | Amber Herold | |
133 | |||
134 | 3 | Amber Herold | * Leginon/Beam Tilt/Measure> click Correct Stigmator to make the correction |
135 | 1 | Amber Herold | according to the measurement. |
136 | |||
137 | |||
138 | * Leginon/Focus/Toolbar> start Manual Focusing and check to see if the |
||
139 | correction makes bad stigmation better. |
||
140 | If not, redo stigmation correction. |
||
141 | |||
142 | |||
143 | |||
144 | Trouble Shooting: |
||
145 | |||
146 | |||
147 | |||
148 | 3 | Amber Herold | * The microscope must be well aligned. The most critical are beam tilt pivot point and rotation center, both should be aligned at eucentric focus. Please refer to [[Microscope Set-up|Microscope Set-up]] to align the microscope well. |
149 | 1 | Amber Herold | |
150 | |||
151 | 3 | Amber Herold | * Chose a beam tilt and defocus change that will give a 10-20% shift in the image. |
152 | 1 | Amber Herold | |
153 | |||
154 | * The beam should cover the entire CCD imaging area. |
||
155 | |||
156 | |||
157 | 3 | Amber Herold | * Have an easily recognized object at the center of the image for easy correlation. Avoid uneven areas. |
158 | 1 | Amber Herold | |
159 | |||
160 | 3 | Amber Herold | * Watch the images movement. There should not be any contrast reversal or appearance of the edge of the beam in any image acquired. Contrast reversal is caused by misaligned objective aperture or by too large a beam tilt angle so that part of the beam is blocked by the aperture when tilted. |
161 | 1 | Amber Herold | |
162 | |||
163 | 3 | Amber Herold | * Check the correlation image and peak location to find out if the image contrast is too low for good correlations. |
164 | 1 | Amber Herold | |
165 | |||
166 | 3 | Amber Herold | * The default values to correct calibration the stagmators are recommended. If these values do not produce good stigmation calibration results and if the microscope has been well aligned, then progressively and slowly change the Stigmator Tilt and Stigmator Delta values until the images cross correlations result in measurable shifts between steps of the calibration. |
167 | 1 | Amber Herold | |
168 | |||
169 | h2. Eucentric Focus Storage and Retrieval |
||
170 | |||
171 | |||
172 | |||
173 | Encentric Focus From/To Scope (From/To Scope icons) do what they indicate. The "from |
||
174 | Scope" function simply records the current absolute focus value at the scope. |
||
175 | Therefore, |
||
176 | |||
177 | |||
178 | 3 | Amber Herold | * Scope> eucenter the grid and reset defocus at U-centric height at the required high tension and at a high mag where focusing is easier. |
179 | 1 | Amber Herold | |
180 | |||
181 | 3 | Amber Herold | * Leginon/Presets Manager> select the preset that will perform z-focusing (i.e., hl). temporarily change its defocus to 0.0 |
182 | 1 | Amber Herold | |
183 | |||
184 | * Leginon/Presets Manager> sent the preset "To Scope" |
||
185 | |||
186 | |||
187 | 3 | Amber Herold | * Leginon/Beam Tilt> left-click the Encentric Focus To Scope icon to save the focus value. |
188 | 1 | Amber Herold | |
189 | |||
190 | * Leginon/Presets Manager> return the defocus value back to its original. |
||
191 | |||
192 | |||
193 | |||
194 | h2. Rotation Center Storage and Retrieval |
||
195 | |||
196 | |||
197 | |||
198 | 3 | Amber Herold | Rotation Center From/To Scope (From/To Scope icons) do what they indicate. The "from Scope" function simply records the current beam tilt value at the scope. Therefore, |
199 | 1 | Amber Herold | |
200 | |||
201 | 3 | Amber Herold | * Scope> Align the rotation center at a magnification and high tension where the value will be used. |
202 | 1 | Amber Herold | |
203 | |||
204 | 3 | Amber Herold | * Leginon/Beam Tilt> left-click the Rotation Center To Scope icon to save the beam tilt value. |
205 | 2 | Amber Herold | |
206 | |||
207 | ______ |
||
208 | |||
209 | [[Checking Matrix and Modeled Stage Position Calibration|< Checking Matrix and Modeled Stage Position Calibration]] | [[Dose (Camera Sensitivity) Calibration|Dose (Camera Sensitivity) Calibration >]] |
||
210 | |||
211 | ______ |