Three functions have to be performed in Beam Tilt node to perform defocus, stigmation and
6
z-height corrections used in autofocus procedures. Defocus and stigmator calibration are
7
required for defocus/stigmation correction. These plus a record of eucentric focus value are
8
required for the z-height correction to eucentric height.
9
10
11 3 Amber Herold
Beam Tilt Defocus/Stigmator Calibration and Eucentric Focus Recording Need for the Example MSI:
12 1 Amber Herold
13 3 Amber Herold
14 1 Amber Herold
|<filename>Preset</filename>|<filename>magnification</filename>|
15
|hl|5000|
16
|fa|50000|
17
18
19
20
21
22
h2. Defocus Calibration
23
24
25
26
27
28 3 Amber Herold
*  Leginon/Presets Managaer> send the preset you want to make the calibration "To Scope"
29 1 Amber Herold
30
31 3 Amber Herold
*  Leginon/Beam Tilt/Toolbar/Setting> check and make sure that it does NOT overwrite presets
32 1 Amber Herold
33
34
*  Leginon/Beam Tilt/Toolbar> select Defocus as calibration type
35
36
37 3 Amber Herold
*  Leginon/Beam Tilt/Toolbar> open "Parameter Settings" window to set parameters as follows:
38 1 Amber Herold
at 50,000x: defocus1 = -2e-06, defocus2 = -4e-06 and beam tilt angle=0.01
39
at 5,000x: defocus1 = -2e-05, defocus2 = -4e-05 and beam tilt angle=0.01
40
41
42
*  Leginon/Beam Tilt/Toolbar> click Calibrate.
43
44
45
*  Occasionally the beam may be partially blocked by the objective aperture and
46
produce a dark field image upon beam tilts. Reducing the beam tilt angle, using a
47
larger objective aperture, and/or centering the objective aperture accurately are the
48
possible solutions.
49
50
51
52
53
h2.  Beam Tilt Stigmator calibration
54
55
56
57
The Beam Tilt Stigmator calibration is used by Leginon to correct astigmatism. This
58
calibration functions well if the microscope has been well aligned first. Stigmation
59
correction at low mag such as 5000x is not necessary practically but the calibration is
60
still required for the equation used in autofocusing.
61
62
63
64
65
*  Leginon/Presets Managaer> if hasn't done so, send the preset you want to make
66
the calibration "To Scope"
67
68
69
*  Leginon/Beam Tilt/Toolbar> select Stigmators for calibration.
70
71
72
*  Leginon/Beam Tilt/Toolbar> open "Parameter Settings" window to set parameters
73
as follows:
74
at 50,000x: beam tilt angle= 0.01 and delta stig.= 0.2
75
at 5,000x: beam tilt angle= 0.01 and delta stig.= 0.2
76
77
78
*  Leginon/Beam Tilt/Toolbar> click Calibrate.
79
80
81
*  Occasionally the beam may be partially blocked by the objective aperture and
82
produce a dark field image upon beam tilts. Reducing the beam tilt angle, using a
83
larger objective aperture, and/or centering the objective aperture accurately are the
84
possible solutions.
85
86
87
88
h2.  Checking the Calibrations
89
90
91
92
Once both Defocus and Stigmation are calibrated against beam-tilt induced image shift,
93
the two can be measured using the Measure Function in the Toolbar.
94
95
96
97
98
*  Leginon/Beam Tilt/Toolbar> click Measure to open measurement submenu.
99
100
101 3 Amber Herold
*  Leginon/Beam Tilt/Measure> click Measure (to test the beam tilt defocus and stigmation calibration).
102 1 Amber Herold
103
104 3 Amber Herold
*  Leginon/Beam Tilt/Measure> click Correct Defocus to make the correction according to the measurement.
105 1 Amber Herold
106
107 3 Amber Herold
*  Leginon/Beam Tilt/Measure> click Reset Defocus and exit the sub-window.
108 1 Amber Herold
109
110
*  Leginon/Presets Managaer> send the fc preset "To Scope"
111
112
113 3 Amber Herold
*  Leginon/Focus/Settings> select fc as the preset used for the acquisition.
114 1 Amber Herold
115
116
*  Leginon/Focus/Toolbar> start Manual Focusing.
117
Power Spectrum of the image at defocus specified by "fc" preset should appear on
118
the screen.
119
120
121
*  Leginon/Focus/Manual Focus> set 0 defocus to scope by selecting the movetype
122
to "defocus" and the number entry in meter to "0" and then click "Set
123
Instrument".
124
The Thone ring should disappear as the defocus is changed to zero measured by Beam
125
Tilt node. If not, recalibration of defocus may be needed.
126
127
128 3 Amber Herold
*  Leginon/Focus/ToolbarManual Focus> stop Manual Focusing by clicking the "stop" button.
129 1 Amber Herold
130
131 3 Amber Herold
*  Leginon/Beam Tilt/Toolbar> click Measure to open measurement sub-menu.
132 1 Amber Herold
133
134 3 Amber Herold
*  Leginon/Beam Tilt/Measure> click Correct Stigmator to make the correction
135 1 Amber Herold
according to the measurement.
136
137
138
*  Leginon/Focus/Toolbar> start Manual Focusing and check to see if the
139
correction makes bad stigmation better.
140
If not, redo stigmation correction.
141
142
143
144
Trouble Shooting:
145
146
147
148 3 Amber Herold
*  The microscope must be well aligned. The most critical are beam tilt pivot point and rotation center, both should be aligned at eucentric focus. Please refer to [[Microscope Set-up|Microscope Set-up]] to align the microscope well.
149 1 Amber Herold
150
151 3 Amber Herold
*  Chose a beam tilt and defocus change that will give a 10-20% shift in the image.
152 1 Amber Herold
153
154
*  The beam should cover the entire CCD imaging area.
155
156
157 3 Amber Herold
*  Have an easily recognized object at the center of the image for easy correlation. Avoid uneven areas.
158 1 Amber Herold
159
160 3 Amber Herold
*  Watch the images movement. There should not be any contrast reversal or appearance of the edge of the beam in any image acquired. Contrast reversal is caused by misaligned objective aperture or by too large a beam tilt angle so that part of the beam is blocked by the aperture when tilted.
161 1 Amber Herold
162
163 3 Amber Herold
*  Check the correlation image and peak location to find out if the image contrast is too low for good correlations.
164 1 Amber Herold
165
166 3 Amber Herold
*  The default values to correct calibration the stagmators are recommended. If these values do not produce good stigmation calibration results and if the microscope has been well aligned, then progressively and slowly change the Stigmator Tilt and Stigmator Delta values until the images cross correlations result in measurable shifts between steps of the calibration.
167 1 Amber Herold
168
169
h2.  Eucentric Focus Storage and Retrieval
170
171
172
173
Encentric Focus From/To Scope (From/To Scope icons) do what they indicate. The "from
174
Scope" function simply records the current absolute focus value at the scope.
175
Therefore,
176
177
178 3 Amber Herold
*  Scope> eucenter the grid and reset defocus at U-centric height at the required high tension and at a high mag where focusing is easier.
179 1 Amber Herold
180
181 3 Amber Herold
*  Leginon/Presets Manager> select the preset that will perform z-focusing (i.e., hl). temporarily change its defocus to 0.0
182 1 Amber Herold
183
184
*  Leginon/Presets Manager> sent the preset "To Scope"
185
186
187 3 Amber Herold
*  Leginon/Beam Tilt> left-click the Encentric Focus To Scope icon to save the focus value.
188 1 Amber Herold
189
190
*  Leginon/Presets Manager> return the defocus value back to its original.
191
192
193
194
h2.  Rotation Center Storage and Retrieval
195
196
197
198 3 Amber Herold
Rotation Center From/To Scope (From/To Scope icons) do what they indicate. The "from Scope" function simply records the current beam tilt value at the scope. Therefore,
199 1 Amber Herold
200
201 3 Amber Herold
*  Scope> Align the rotation center at a magnification and high tension where the value will be used.
202 1 Amber Herold
203
204 3 Amber Herold
*  Leginon/Beam Tilt> left-click the Rotation Center To Scope icon to save the beam tilt value.
205 2 Amber Herold
206