h1. Autofocus Calibration with Beam Tilt Node
2
3
4
5
Three functions have to be performed in Beam Tilt node to perform defocus, stigmation and
6
z-height corrections used in autofocus procedures. Defocus and stigmator calibration are
7
required for defocus/stigmation correction. These plus a record of eucentric focus value are
8
required for the z-height correction to eucentric height.
9
10
11 3 Amber Herold
Beam Tilt Defocus/Stigmator Calibration and Eucentric Focus Recording Need for the Example MSI:
12 1 Amber Herold
13 3 Amber Herold
14 1 Amber Herold
|<filename>Preset</filename>|<filename>magnification</filename>|
15
|hl|5000|
16
|fa|50000|
17
18
19
20
21
22
h2. Defocus Calibration
23
24
25 4 Amber Herold
#  Leginon/Presets Managaer> send the preset you want to make the calibration "To Scope"
26
&nbsp;
27
#  Leginon/Beam Tilt/Toolbar/Setting> check and make sure that it does NOT overwrite presets
28
&nbsp;
29
#  Leginon/Beam Tilt/Toolbar> select Defocus as calibration type
30
&nbsp;
31
#  Leginon/Beam Tilt/Toolbar> open "Parameter Settings" window to set parameters as follows:
32
&nbsp;
33 1 Amber Herold
at 50,000x: defocus1 = -2e-06, defocus2 = -4e-06 and beam tilt angle=0.01
34 4 Amber Herold
&nbsp;
35 1 Amber Herold
at 5,000x: defocus1 = -2e-05, defocus2 = -4e-05 and beam tilt angle=0.01
36 4 Amber Herold
&nbsp;
37
#  Leginon/Beam Tilt/Toolbar> click Calibrate.
38
&nbsp;
39
#  Occasionally the beam may be partially blocked by the objective aperture and
40 1 Amber Herold
produce a dark field image upon beam tilts. Reducing the beam tilt angle, using a
41
larger objective aperture, and/or centering the objective aperture accurately are the
42
possible solutions.
43
44
45
h2.  Beam Tilt Stigmator calibration
46
47
48
49
The Beam Tilt Stigmator calibration is used by Leginon to correct astigmatism. This
50
calibration functions well if the microscope has been well aligned first. Stigmation
51
correction at low mag such as 5000x is not necessary practically but the calibration is
52
still required for the equation used in autofocusing.
53
54
55
56
57 4 Amber Herold
#  Leginon/Presets Managaer> if hasn't done so, send the preset you want to make
58 1 Amber Herold
the calibration "To Scope"
59 4 Amber Herold
&nbsp;
60
#  Leginon/Beam Tilt/Toolbar> select Stigmators for calibration.
61
&nbsp;
62
#  Leginon/Beam Tilt/Toolbar> open "Parameter Settings" window to set parameters
63 1 Amber Herold
as follows:
64
at 50,000x: beam tilt angle= 0.01 and delta stig.= 0.2
65
at 5,000x: beam tilt angle= 0.01 and delta stig.= 0.2
66 4 Amber Herold
&nbsp;
67
#  Leginon/Beam Tilt/Toolbar> click Calibrate.
68
&nbsp;
69
#  Occasionally the beam may be partially blocked by the objective aperture and
70 1 Amber Herold
produce a dark field image upon beam tilts. Reducing the beam tilt angle, using a
71
larger objective aperture, and/or centering the objective aperture accurately are the
72
possible solutions.
73
74
75
76
h2.  Checking the Calibrations
77
78
79
80 3 Amber Herold
Once both Defocus and Stigmation are calibrated against beam-tilt induced image shift,
81 1 Amber Herold
the two can be measured using the Measure Function in the Toolbar.
82
83 3 Amber Herold
84 1 Amber Herold
85
86 4 Amber Herold
#  Leginon/Beam Tilt/Toolbar> click Measure to open measurement submenu.
87
&nbsp;
88
#  Leginon/Beam Tilt/Measure> click Measure (to test the beam tilt defocus and stigmation calibration).
89
&nbsp;
90
#  Leginon/Beam Tilt/Measure> click Correct Defocus to make the correction according to the measurement.
91
&nbsp;
92
#  Leginon/Beam Tilt/Measure> click Reset Defocus and exit the sub-window.
93
&nbsp;
94
#  Leginon/Presets Managaer> send the fc preset "To Scope"
95
&nbsp;
96
#  Leginon/Focus/Settings> select fc as the preset used for the acquisition.
97
&nbsp;
98
#  Leginon/Focus/Toolbar> start Manual Focusing.
99
&nbsp;
100 1 Amber Herold
Power Spectrum of the image at defocus specified by "fc" preset should appear on
101 3 Amber Herold
the screen.
102 4 Amber Herold
&nbsp;
103
#  Leginon/Focus/Manual Focus> set 0 defocus to scope by selecting the movetype
104 1 Amber Herold
to "defocus" and the number entry in meter to "0" and then click "Set
105
Instrument".
106 4 Amber Herold
&nbsp;
107 1 Amber Herold
The Thone ring should disappear as the defocus is changed to zero measured by Beam
108
Tilt node. If not, recalibration of defocus may be needed.
109 4 Amber Herold
&nbsp;
110
#  Leginon/Focus/ToolbarManual Focus> stop Manual Focusing by clicking the "stop" button.
111
&nbsp;
112
#  Leginon/Beam Tilt/Toolbar> click Measure to open measurement sub-menu.
113
&nbsp;
114
#  Leginon/Beam Tilt/Measure> click Correct Stigmator to make the correction
115 1 Amber Herold
according to the measurement.
116 4 Amber Herold
&nbsp;
117
#  Leginon/Focus/Toolbar> start Manual Focusing and check to see if the
118 1 Amber Herold
correction makes bad stigmation better.
119
If not, redo stigmation correction.
120
121
122 3 Amber Herold
123 1 Amber Herold
Trouble Shooting:
124
125 3 Amber Herold
126 1 Amber Herold
127
*  The microscope must be well aligned. The most critical are beam tilt pivot point and rotation center, both should be aligned at eucentric focus. Please refer to [[Microscope Set-up|Microscope Set-up]] to align the microscope well.
128
129
130
*  Chose a beam tilt and defocus change that will give a 10-20% shift in the image.
131 3 Amber Herold
132 1 Amber Herold
133
*  The beam should cover the entire CCD imaging area.
134 3 Amber Herold
135 1 Amber Herold
136
*  Have an easily recognized object at the center of the image for easy correlation. Avoid uneven areas.
137 3 Amber Herold
138 1 Amber Herold
139
*  Watch the images movement. There should not be any contrast reversal or appearance of the edge of the beam in any image acquired. Contrast reversal is caused by misaligned objective aperture or by too large a beam tilt angle so that part of the beam is blocked by the aperture when tilted.
140 3 Amber Herold
141 1 Amber Herold
142
*  Check the correlation image and peak location to find out if the image contrast is too low for good correlations.
143
144
145
*  The default values to correct calibration the stagmators are recommended. If these values do not produce good stigmation calibration results and if the microscope has been well aligned, then progressively and slowly change the Stigmator Tilt and Stigmator Delta values until the images cross correlations result in measurable shifts between steps of the calibration.
146
147
148
h2.  Eucentric Focus Storage and Retrieval
149
150
151
152 3 Amber Herold
Encentric Focus From/To Scope (From/To Scope icons) do what they indicate. The "from
153 1 Amber Herold
Scope" function simply records the current absolute focus value at the scope.
154
Therefore,
155 3 Amber Herold
156 1 Amber Herold
157 4 Amber Herold
#  Scope> eucenter the grid and reset defocus at U-centric height at the required high tension and at a high mag where focusing is easier.
158
&nbsp;
159
#  Leginon/Presets Manager> select the preset that will perform z-focusing (i.e., hl). temporarily change its defocus to 0.0
160
&nbsp;
161
#  Leginon/Presets Manager> sent the preset "To Scope"
162
&nbsp;
163
#  Leginon/Beam Tilt> left-click the Encentric Focus To Scope icon to save the focus value.
164
&nbsp;
165
#  Leginon/Presets Manager> return the defocus value back to its original.
166 1 Amber Herold
167
168
169
h2.  Rotation Center Storage and Retrieval
170
171
172 3 Amber Herold
173 1 Amber Herold
Rotation Center From/To Scope (From/To Scope icons) do what they indicate. The "from Scope" function simply records the current beam tilt value at the scope. Therefore,
174
175 3 Amber Herold
176 4 Amber Herold
#  Scope> Align the rotation center at a magnification and high tension where the value will be used.
177
&nbsp;
178
#  Leginon/Beam Tilt> left-click the Rotation Center To Scope icon to save the beam tilt value.
179 2 Amber Herold
180
181
______
182
183
[[Checking Matrix and Modeled Stage Position Calibration|< Checking Matrix and Modeled Stage Position Calibration]] | [[Dose (Camera Sensitivity) Calibration|Dose (Camera Sensitivity) Calibration >]]
184
185
______