Project

General

Profile

Bright and Dark reference images » History » Version 10

Amber Herold, 06/23/2010 12:48 PM

1 1 Amber Herold
h1. Bright and Dark reference images
2
3
4
Bright and Dark reference images need to be acquired for every camera setting that will
5
be used. The camera settings include image dimension, bin size, and offset. Over time,
6
references may need to be repeatedly acquired.
7
8
9
10
h2. Correction Channels
11
12
13
14
When two flat-field-corrected images are correlated, there is often an origin peak
15
derived from the common normalization image even if both image acquisition contains only
16
noise. In order to avoid this problem, two or more sets of bright/dark references, and hence
17
normalization images can be obtained per CCD camera configuration. When a correlation
18
between two images will be done, Leginon will check the channel of the correction the first
19
acquired image has used and then force the new image to be corrected by a different
20
channel.
21
22
23
To use this function, simple set the number of channels to 2 or larger in
24
Correction/Settings/Image Correction/Reference Creation> while creating the reference
25
images.
26
27
28
29
30
31
h2. Acquire reference images
32
33
34
35
*  scope> make sure that the CCD will be acquiring images in an area with uniform
36
beam intensity such as an empty area with no specimen nor support. You may skip a trip
37
to the scope room by sending one of the high mag preset to the scope from
38
Leginon.
39
40
41
*  Leginon/Node Selector> Select "Correction" node.
42
43
44
*  Leginon/Correction/Toolbar> Open "Settings" window.
45
46
47
*  Leginon/Correction/Toolbar/Settings> Select one of the Common Camera
48
Configuration or select Custom mode and enter your own values based on the presets you
49
created.
50
51
52
*  Leginon/Correction/Settings/Camera Configuration> Enter the Exposure time. It
53
should be chosen so that the image is not saturated and ideally close to the condition
54
that will be used in the experiments. If unsure about the experimental condition, use an
55
exposure time that gives high but not saturated counts.
56
57
58
*  Leginon/Correction/Settings>By default, the corrector node is set to average 3
59
images together to create one reference image and to despike the hot pixels with
60
averaged neighbor hood values. These can be changed if desired.
61
62
63 10 Amber Herold
*  Leginon/Correction/Settings> It is recommended to use 2 [[Bright and Dark reference images#Correction-Channels|correction channels]] for the camera configuration used
64 1 Amber Herold
for correlation peak search.
65
66
67
*  Leginon/Correction/Settings> Click OK to exit settings.
68
69
70
*  Leginon/Correction/Toolbar> Select "Raw image " from the pull down list of
71
acquisition modes and then click on "Acquire" button next to the selector to view an
72
image that is not corrected.
73
74
75
*  Leginon/Correction/Toolbar> Select "Dark reference" in the acquisition mode
76
selector and then click "Acquire" to acquire the Dark reference image for this
77
particular camera configuration.
78
79
80
*  Leginon/Correction/Toolbar> Select "Bright reference" and repeat the acquisition
81
to obtain the Bright reference.
82
83
84
*  Leginon/Correction/Toolbar> Select "Corrected image" and then "Acquire" to view
85
the corrected image. A corrected image should be free of artifacts and have smaller
86
standard deviation than the raw image, in general.
87
88
89
*  Repeat steps 3-11 for all the images and bin sizes that will be used:
90
91
92 6 Alex Kadokura
*  If [[Bright and Dark reference images#Image-Despike|a pixel, a column/row]] or a [[Bright and Dark reference images#Image-Despike|region]] gives bad values in the bright or dark image
93 1 Amber Herold
after a few trials, it may be excluded in all corrected images.
94
95 7 Anchi Cheng
* Only one channel is required for image shift alignment purpose. If the same camera configuration is used for presets that involves in correlation, two channels are strongly recommended.
96 1 Amber Herold
97 3 Anchi Cheng
|*Dimension after binning*|*Bin*|*number of correction channels*|
98 1 Amber Herold
|4096|1|1 or 2 if used for tomo preset|
99 9 Anchi Cheng
|1024|4|2^*^|
100
|1024(centered)|1|1^**^|
101
|512|8|2^***^|
102
|512 (centered)|1|1^***^|
103 1 Amber Herold
104 9 Anchi Cheng
^*^-This camera configuration is used in preset beam shift alignment even if you don't use it for a preset.
105
^**^-This camera configuration is used in Manual Application Manual Focusing even if you don't use it for a preset.
106
^***^-These camera configurations are used in preset image shift alignment even if you don't use it for a preset.
107 1 Amber Herold
108
h2. Image Despike
109
110
111
112
The Despike feature removes random bright or hot pixels from the acquired images. This
113
hot pixel is assigned the average intensity of the surrounding area, a circle of the radius
114
which is entered in Neighborhood Size. The Despike Threshold is the number of standard
115
deviations away from the mean that qualifies a pixel for despike correction. The despike
116
affects the flat-field corrected image saved on the disk and can not be recovered.
117
Therefore, use a minimal neighborhood size to avoid artifact and set the threshold high to
118
avoid over-dispiking.
119
120 3 Anchi Cheng
Activation of this feature and its parameter settings are defined when in the pop-up dialog for "Edit Correction Plan".  See below.
121 1 Amber Herold
122
123
124
125 3 Anchi Cheng
126 1 Amber Herold
h2. Correction Plan
127
128
129
130
Bad Pixel, Rows and Bad Cols are used to crop portions of the image that do not read
131
well off of the CCD. The values entered into here are determined empirically for each
132
instrument that Leginon operates on. If one column or row of the images is incorrect,
133
measure the location of the row and column that need to be removed from this image. These
134
values should then be entered as a sequence of values separated by commas by editing the
135
Plan. Click Save after adjusting.
136
137
138
Individual bad pixel can also be corrected by its surrounding pixels. Choose these
139
pixels with the selection tool on the image and then click on "Grab From Image".
140
141
142
143
144
145
h2. Find A Single Bad Pixel
146
147
148
149
When a single pixel is defected, it may not be easy to find it on a large image, even if
150
it changes the stats dramatically. A tool is available to help finding these pixels:
151
152
153
*  Leginon/Correction> Acquire either a corrected image that shows the bad
154
stats.
155
156
157 3 Anchi Cheng
*  Leginon/Correction/Toolbar> Left-click on the !http://emg.nysbc.org/software/leginon/images/icons/stagelocations.png!
158 1 Amber Herold
button to "Add extreme points to bad pixel list". There
159
160
161
*  Leginon/Correction/Tools> Left-click on the "Add Region" tool that looks like
162
"+". This adds the selected bad region to the bad pixel plan.
163
164
165
*  Leginon/Corrections> Acquire a corrected image in the same configuration to
166
check if the apearance improves.
167
168
169
170
171
172
173
174
h2. Bad Region Correction
175
176
177
178
When a large region is covered by a fallen chip, image correction through bright/dark
179
reference may not be sufficient to produce a spike-free image since the bright and dark
180
values in the region are almost identical. To add such a large region into bad pixel plan,
181
do the following:
182
183
184
*  Leginon/Correction> Acquire either a bright or corrected image that shows the
185
bad region clearly.
186
187
188
*  Leginon/Correction> Use "Regions" target tool next to the image to enclose the
189
bad region. The corners that the target tool identifies can be larger than the bad
190
region but should be close to its size so that not too much is corrected.
191
192
193
*  Leginon/Correction/Tools> Left-click on the "Add Region" tool that looks like
194
"+". This adds the selected bad region to the bad pixel plan.
195
196
197
*  Leginon/Corrections> Acquire a corrected image in the same configuration to
198 10 Amber Herold
check if the appearance improves.
199 2 Amber Herold
200
______
201
202
[[Pixel Size Calibration|< Pixel Size Calibration]] | [[Image Shift matrix calibration|Image Shift matrix calibration >]]
203
204
______