1 1 Anchi Cheng
h1. Camera length calibration
2
3
Camera length calibration can be simply done by acquiring a diffraction image of cross-grating replica and enter the expected d-spacing of Au-Pd power rings and measure the diameter of the ring in Leginon gui.
4
5
# Send a diffraction tem preset. *This should be a square image*
6
# Select "Camera Length" node
7
# Click "Acquire Image"
8
# Measure the diameter of the ring in pixels.
9
# Click "Ruler" tool to enter the expected d-spacing and measurement, and then click "Calculate"
10
# Multiple measurements can be made with the same preset and the camera length averaged.
11
# Select the camera lengths to be averaged and saved, and save them.
12
13
Direct Edit and Extrapolation is possible as in Pixel Size Calibrator.