Camera length calibration » History » Version 1
  Anchi Cheng, 02/08/2020 11:12 PM 
  
| 1 | 1 | Anchi Cheng | h1. Camera length calibration | 
|---|---|---|---|
| 2 | |||
| 3 | Camera length calibration can be simply done by acquiring a diffraction image of cross-grating replica and enter the expected d-spacing of Au-Pd power rings and measure the diameter of the ring in Leginon gui. | ||
| 4 | |||
| 5 | # Send a diffraction tem preset. *This should be a square image* | ||
| 6 | # Select "Camera Length" node | ||
| 7 | # Click "Acquire Image" | ||
| 8 | # Measure the diameter of the ring in pixels. | ||
| 9 | # Click "Ruler" tool to enter the expected d-spacing and measurement, and then click "Calculate" | ||
| 10 | # Multiple measurements can be made with the same preset and the camera length averaged. | ||
| 11 | # Select the camera lengths to be averaged and saved, and save them. | ||
| 12 | |||
| 13 | Direct Edit and Extrapolation is possible as in Pixel Size Calibrator. | ||
| 14 |