Camera length calibration¶
Camera length calibration can be simply done by acquiring a diffraction image of cross-grating replica and enter the expected d-spacing of Au-Pd power rings and measure the diameter of the ring in Leginon gui.
- This is only tested on TFS Ceta and Tvips F416 at 2kx2k bin 2
- Send a diffraction tem preset. This should be a square image.
- Select "Camera Length" node
- Click "Acquire Image"
- Measure the diameter of the ring in pixels.
- Click "Ruler" tool to enter the expected d-spacing and measurement, and then click "Calculate"
- Multiple measurements can be made with the same preset and the camera length averaged.
- Select the camera lengths to be averaged and saved, and save them.
Direct Edit and Extrapolation is possible as in Pixel Size Calibrator.
Updated by Anchi Cheng over 4 years ago · 2 revisions