h1. Camera length calibration 

 Camera length calibration can be simply done by acquiring a diffraction image of cross-grating replica and enter the expected d-spacing of Au-Pd power rings and measure the diameter of the ring in Leginon gui. 

 It will also save a pixel size equivalent that allows the webviewer to display the diffraction image and use its ruler to estimate the resolution. 

 * This is only tested on TFS Ceta and Tvips F416 at 2kx2k bin 2 
 
 # Send a diffraction tem preset.    See [[Setup df preset]] 
 # Select "Camera Length" node 
 # Click "Acquire Image" 
 # Measure the diameter of the ring in pixels. 
 # Click "Ruler" tool to enter the expected d-spacing and measurement, and then click "Calculate" 
 # Multiple measurements can be made with the same preset and the camera length averaged. 
 # Select the camera lengths to be averaged and saved, and save them. 

 Direct Edit and Extrapolation is possible just like as in Pixel Size Calibrator.