Actions
Continuous tilt electron diffraction recording » History » Revision 3
« Previous |
Revision 3/37
(diff)
| Next »
Anchi Cheng, 12/05/2019 07:44 PM
Continuous tilt electron diffraction recording¶
Micro-ED involves recording 3D crystal electron diffraction pattern during a continuous stage tilt. Leginon implementation of this application includes the listed components below and is available by using myami-ed branch until it is officially added to future release. As is, it is only fully implemented for TFS Ceta or Ceta-D camera.
Since many of the feature required here are not available through Standard or Advanced TEM Scripting from TFS, AutoIt scripts are used in several places.
Components of the feature¶
- Diffraction mode TEM instrument
- Beamstop control
- Rolling-shutter movie acquisition using TIA interface
- Data conversion and upload of the movies into Leginon database and crystallography format
- MSI-diffraction application
Installation (assuming that you've already have Leginon installation).¶
- Check that you have access to Advanced TEM Scripting (frame-saving upgrade not needed).
- Install or set environment to use git branch myami-ed on TFS microscope with Ceta camera as well as your leginon linux box.
- AutoIt program and script compilation
- import MSI-diffraction application and settings from your_myami/leginon/applications (see Steps_involved_in_the_installation regarding importing application and additional settings for the application)
Usage¶
(To be written)