Project

General

Profile

Create Leginon view port » History » Version 1

Anchi Cheng, 07/29/2021 02:07 PM

1 1 Anchi Cheng
h1. Create Leginon view port
2
3
EM-Menu use saved and loaded viewports to set preset configuration.  In order not to disturb its other behavior, Leginon
4
uses a specific named viewport when it controls the camera.
5
6
1. Open EM-Menu on TVIPS computer.  The last configuration and view port used is typically appeared and activated.
7
8
2. Select from menubar Camera/Configuration Manager ...
9
10
3. Click New to create a configuration, name it Leginon
11
12
Here is an example:
13
14
* Format
15
!viewport_format.png!
16
17
* Shutter
18
!viewport_shutter.png!
19
20
* Flatfield
21
!viewport_flatfield.png!
22
23
4. Save the current configuration list as a backup.
24
25
5. close the dialog and reopen by repeating step 2.  Leginon should either be selected or at least listed as an existing configuration.
26
27
h2. Testing
28
29
1. Close EM-Menu and then reopen.  Use step 2 in the previous section to check if the new viewport is loaded.