Project

General

Profile

Creating Presets for the first time » History » Revision 2

Revision 1 (Amber Herold, 04/22/2010 04:49 PM) → Revision 2/10 (Amber Herold, 04/27/2010 10:42 AM)

h1. Creating Presets for the first time 



 Configuring the Presets are critical to many different applications. Presets describe 
 camera configurations such as image size and exposure time as well as many other microscope 
 parameters such as magnification and defocus. Setting presets in calibration application is 
 optional but recommended for determining what calibrations are needed for future runs. 



 h2. Design Presets 



 MSI presets can be customized to your need of data collection. By matching 
 nodes/subnodes in MSI with presets and move types, the behavior of the node is 
 defined. 


 See the section on <link linkend="preset_design">Designing Presets</link> in the chapter 
 for MSI application for more explanations on the presets standardized at NRAMM which we will 
 reference in this chapter, we use here the nomenclatures at NRAMM that is used to collect 
 defocus paired images with four scales of targeting ant two sequences of focusing using an 
 MSI application. You will need to adjust the preset parameters for your own camera, scope, 
 as well as the grid mesh you use. 





 h2. Presets Created for Convenience in Calibration 



 The table shows an example of the presets for an MSI application. For calibration 
 purpose, undefined image shift can be left as what recieved from the scope. Beam intensity, 
 spot size, and should be adjusted to give reasonable signal to noise ratio for usable 
 exposure time and for the beam to cover the whole CCD. The exact setting of exposure time 
 can be set later after bright and dark images are collected. 



 |_.Magnification:|_.Preset name:|_.Image Shift (x,y):|_.Binning:|_.Dimension:|_.Beam Coverage:|_.Exposure Time (ms):|_.Spot Size:|_.Defocus (m):| 
 |120|gr|as is*|8|512|max|20|4|0.0| 
 |550|sq|as is*|4|1024|1x CCD size|100|4|-2e-3| 
 |5000|hl|as is*|8|512|1x CCD|20|4|-1.5e-4| 
 |50000|fc|0,0|1|512|<~ 1x CCD|300|4|-2e-6| 
 |50000|fa|0,0|4|1024|>2x CCD|50|4|-2e-6| 
 |50000|en|0,0|1|4096|2x CCD|170 (10e/A^2)|4|-1e-6| 
 |50000|ef|0,0|1|4096|2x CCD|170 (10e/A^2)|4|-2e-6| 




 * Use whatever image shift from the microscope when the preset is newly created. These 
 will be aligned later. 





 h2. Create Presets from microscope setting 





 *    (Optional) Microscope/Reset defocus at eucentric height. 


 *    Adjust the microscope settings at the microscope. This includes magnification, 
 beam intensity, beam shift, and defocus. Beam intensity and beam shift is adjusted to 
 get the good beam coverage shown in the table. Defocus only need to be 
 approximate. 


 *    Leginon/Node Selector>Select "Presets Manager" node. 


 *    Leginon/Presets Manager> Click New From Scope icon !http://emg.nysbc.org/software/leginon/images/icons/instrumentgetnew.png! to open "Create New Preset" setting window. 


 *    Leginon/Presets Manager/Create New Preset> Type a descriptive name for a 
 Preset, "hl" for example, left-click "Create" to save and close the window. 


 *    Leginon/Presets Manager/Preset Parameters/Camera Configuration> Change the 
 image shift, defocus, camera configuration and exposure time if necessary. It will be 
 saved automatically when changed. 









 h2. Presets Selector and Cycle Order List 



 If Presets are not ordered as in the table above, we recommend that they are rearranged 
 for efficient cycling. Cycling is important for minimizing hysteresis effect on em optics 
 and targeting and are defaulted to be on in Presets Manager Settings. To reduce cycling 
 time, it is best to group the presets that often occur consecutively in order. For example, 
 the typical MSI preset order as in the table above. 




 *    Leginon/Presets Manager> select a preset from the Preset Selector. 


 *    Up/Down: Move the presets up or down in the Cycle Order. 









 h2. Adjusting Presets at the microscope 



 With cycling activated as default, the preset beam shift may behave differently from 
 that was first created. The following procedure check, adjust, and save the presets 
 parameters with cycling accounted for. 




 *    Leginon/Presets Manager> select a preset from the Preset Selector. 


 *    Leginon/Presets Manager> click To Scope icon !http://emg.nysbc.org/software/leginon/images/icons/instrumentset.png! to send the current preset to the microscope. 


 *    Adjust the microscope settings either at the microscope or in a node that 
 controls the microscope (Instrument) and/or the camera (Navigation,etc.). 


 *    Leginon/Presets Manager> click From Scope icon !http://emg.nysbc.org/software/leginon/images/icons/instrumentget.png! to save the current settings to this preset. 









 h2. Using Presets in Calibrations 



 With presets created, calibrations at different magnifications can be changed by sending 
 the corresponding preset to the scope. 




 *    Leginon/Presets Manager> select a preset from the Preset Selector. 


 *    Leginon/Presets Manager> click To Scope icon to send the current preset to the 
 microscope. 









 h2. Calibration Status 


 Calibration status of a preset is displayed in PresetsManager right below the message 
 log panel. This section is very useful for troubleshooting calibration problems. Over time, 
 certain calibrations may need to be repeated. This section will display the latest 
 calibration date related to the current preset. 




 Check this section to see which calibrations the new preset will need to be 
 completed. 




 *    Pixel Size: This value is entered through the Pixel Size Calibrator node. 
 (required) 


 *    Image Shift: This is calculated through the Image Shift matrix calibration. 
 (required) 


 *    Stage: This is calculated through the Stage Shift matrix calibration. (Optional 
 if Modeled Stage Mag Only calibraion exists at the same mag). 


 *    Beam Shift: This is calculated through the Beam Shift matrix calibration 
 (Optional; Necessary to move the beam in Navigator). 


 *    Modeled Stage: This is calculated through the Modeled Stage Position calibration 
 (necessary to accurately center holes/targets for high magnification images). 


 *    Modeled Stage Mag Only: Same function as in 5. This may be created from full or 
 mag only model fit in GonioModeling. 



 ______ 

 [[Startup Calibrations|< Startup]] | [[Calibration without presets|Calibration without presets >]] 

 ______