h1. Dose (Camera Sensitivity) Calibration
2
3
4
5
Dose Calibration is a calibration of CCD sensitivity. This calibration is required only
6
once at each high tension per instrument. It can be used for three purposes:
7
8
9
10
11
*  To obtain a scaling factor between the scope screen current and the beam current
12
measured by a standardized equipment such as a Faraday cup.
13
14
15
*  To calibrate the camera sensitivity to electron.
16
17
18
*  If camera sensitivity is already calibrated, the value can be entered
19
directly.
20
21
22
23
24
25
26
27
h2.  Determine the screen current/beam current scale factor
28
29
30
31
The following uses a Faraday cup
32
33
34
35
36
*  Scope> Move the stage to an empty area or a broken hole.
37
38
39
*  Scope> Adjust the beam so that it passes through the measuring area of the
40
Faraday cup.
41
42
43
*  Scope> Choose a magnification so that the same beam will pass within the main
44
screen, too.
45
46
47
*  Leginon/Dose/Dose Calibrator> Beam Diameter = 0.16 m for Tecnai main
48
screen.
49
50
51
*  Leginon/Dose/Dose Calibrator> Enter Screen Current to Beam Current Scale
52
Factor as 1 as a starting point.
53
54
55
*  scope> Optionally lower the screen. Dose Calibrator will automatically do this
56
if the user forgets.
57
58
59
*  Leginon/Dose/Dose Calibrator> Click Measure Dose Rate. The Dose Rate, Beam
60
Current, and Screen Magnification will appear in the corresponding fields after they
61
have been calculated.
62
63
64
*  Leginon/Dose/Dose Calibrator> Adjust the Screen Current to Beam Current Scale
65
Factor so that the displayed beam current is equal to that of the Faraday cup
66
measurement. NRAMM Tecnai's value is 0.88.
67
68
69
70
71
72
73
74
75
76
h2.  Calibrate the camera sensitivity
77
78
79
80
81
82
*  Scope> Move the stage to an empty area or a broken hole.
83
84
85
*  Scope> Expand the beam so that it covers the main screen and illuminate it
86
uniformly.
87
88
89
*  Leginon/Dose/Settings> Configure camera with overide preset activated.
90
91
92
*  Leginon/Dose/Dose Calibrator>
93
94
95
96
Beam Diameter = 0.16 m for the Tecnai's main viewing screen.
97
98
99
Screen Current to Beam Current Scale Factor = (as determined above or 1 to assume
100
good calibration by the microscope manufacturer)
101
102
103
*  scope> Optionally lower the screen. Dose Calibrator will automatically do this
104
if the user forgets.
105
106
107
*  Leginon/Dose/Dose Calibrator> Click Measure Dose Rate. The Dose Rate, Beam
108
Current, and Screen Magnification will appear in the corresponding fields after they
109
have been calculated.
110
111
112
*  Leginon/Dose/Dose Calibrator> Without changing the beam intensity, click on
113
"Calibrate". An image is acquired by Leginon.
114
115
116
*Leginon will automatically lift the main viewing screen on the microscope. Some
117
users report that the function does not work on their microscopes. Do it manually if
118
the automatic function does not work.
119
120
121
*  The calculated CCD sensitivity (counts / electron) will display once the
122
calculation is completed.
123
124
125
126
127
128
129
130
131
132
h2.  Enter pre-determined camera sensitivity value
133
134
135
136
137
138
*  Leginon/Dose/Dose Settings/Camera Sensitivity> Enter Sensitivity values in
139
counts per electron.
140
141
142
*  Leginon/Dose/Dose Settings/Camera Sensitivity> Click on Save.
143
144
145
146
147
148
149
150
|_.Preset|_.magnification|
151
|*|any|