Dose Calibration is a calibration of CCD sensitivity. This calibration is required only once at each high tension per instrument. It can be used for three purposes:
6 1 Amber Herold
7
8 3 Amber Herold
*  To obtain a scaling factor between the scope screen current and the beam current measured by a standardized equipment such as a Faraday cup.
9 1 Amber Herold
10
11
*  To calibrate the camera sensitivity to electron.
12
13
14 3 Amber Herold
*  If camera sensitivity is already calibrated, the value can be entered directly.
15 1 Amber Herold
16
17
18
h2.  Determine the screen current/beam current scale factor
19
20
21
22
The following uses a Faraday cup
23
24
25
26
*  Scope> Move the stage to an empty area or a broken hole.
27
28
29 3 Amber Herold
*  Scope> Adjust the beam so that it passes through the measuring area of the Faraday cup.
30 1 Amber Herold
31
32 3 Amber Herold
*  Scope> Choose a magnification so that the same beam will pass within the main screen, too.
33 1 Amber Herold
34
35 3 Amber Herold
*  Leginon/Dose/Dose Calibrator> Beam Diameter = 0.16 m for Tecnai main screen.
36 1 Amber Herold
37
38 3 Amber Herold
*  Leginon/Dose/Dose Calibrator> Enter Screen Current to Beam Current Scale Factor as 1 as a starting point.
39 1 Amber Herold
40
41 3 Amber Herold
*  scope> Optionally lower the screen. Dose Calibrator will automatically do this if the user forgets.
42 1 Amber Herold
43
44 3 Amber Herold
*  Leginon/Dose/Dose Calibrator> Click Measure Dose Rate. The Dose Rate, Beam Current, and Screen Magnification will appear in the corresponding fields after they have been calculated.
45 1 Amber Herold
46
47 3 Amber Herold
*  Leginon/Dose/Dose Calibrator> Adjust the Screen Current to Beam Current Scale Factor so that the displayed beam current is equal to that of the Faraday cup measurement. NRAMM Tecnai's value is 0.88.
48 1 Amber Herold
49
50
51
h2.  Calibrate the camera sensitivity
52
53
54
55
*  Scope> Move the stage to an empty area or a broken hole.
56
57
58 3 Amber Herold
*  Scope> Expand the beam so that it covers the main screen and illuminate it uniformly.
59 1 Amber Herold
60
61 3 Amber Herold
*  Leginon/Dose/Settings> Configure camera with override preset activated.
62 1 Amber Herold
63
64
*  Leginon/Dose/Dose Calibrator>
65 3 Amber Herold
** Beam Diameter = 0.16 m for the Tecnai's main viewing screen.
66
** Screen Current to Beam Current Scale Factor = (as determined above or 1 to assume
67 1 Amber Herold
good calibration by the microscope manufacturer)
68
69
70 3 Amber Herold
*  scope> Optionally lower the screen. Dose Calibrator will automatically do this if the user forgets.
71 1 Amber Herold
72
73 3 Amber Herold
*  Leginon/Dose/Dose Calibrator> Click Measure Dose Rate. The Dose Rate, Beam Current, and Screen Magnification will appear in the corresponding fields after they have been calculated.
74 1 Amber Herold
75
76 3 Amber Herold
*  Leginon/Dose/Dose Calibrator> Without changing the beam intensity, click on "Calibrate". An image is acquired by Leginon.
77
*Leginon will automatically lift the main viewing screen on the microscope. Some users report that the function does not work on their microscopes. Do it manually if the automatic function does not work.
78 1 Amber Herold
79
80 3 Amber Herold
*  The calculated CCD sensitivity (counts / electron) will display once the calculation is completed.
81 1 Amber Herold
82
83
84
85
h2.  Enter pre-determined camera sensitivity value
86
87
88 3 Amber Herold
*  Leginon/Dose/Dose Settings/Camera Sensitivity> Enter Sensitivity values in counts per electron.
89 1 Amber Herold
90
91
*  Leginon/Dose/Dose Settings/Camera Sensitivity> Click on Save.
92
93 3 Amber Herold
Dose Calibration Need for the Example MSI: