Dose (Camera Sensitivity) Calibration » History » Version 7
Jim Pulokas, 10/04/2010 03:11 PM
1 | 1 | Amber Herold | h1. Dose (Camera Sensitivity) Calibration |
---|---|---|---|
2 | |||
3 | 7 | Jim Pulokas | [[Faraday Cup Measurment Notes]] |
4 | 1 | Amber Herold | |
5 | 3 | Amber Herold | Dose Calibration is a calibration of CCD sensitivity. This calibration is required only once at each high tension per instrument. It can be used for three purposes: |
6 | 1 | Amber Herold | |
7 | 4 | Amber Herold | # To obtain a scaling factor between the scope screen current and the beam current measured by a standardized equipment such as a Faraday cup. |
8 | |
||
9 | # To calibrate the camera sensitivity to electron. |
||
10 | |
||
11 | # If camera sensitivity is already calibrated, the value can be entered directly. |
||
12 | 1 | Amber Herold | |
13 | |||
14 | |||
15 | h2. Determine the screen current/beam current scale factor |
||
16 | |||
17 | |||
18 | |||
19 | The following uses a Faraday cup |
||
20 | |||
21 | |||
22 | |||
23 | 4 | Amber Herold | # Scope> Move the stage to an empty area or a broken hole. |
24 | |
||
25 | # Scope> Adjust the beam so that it passes through the measuring area of the Faraday cup. |
||
26 | |
||
27 | # Scope> Choose a magnification so that the same beam will pass within the main screen, too. |
||
28 | |
||
29 | # Leginon/Dose/Dose Calibrator> Beam Diameter = 0.16 m for Tecnai main screen. |
||
30 | |
||
31 | # Leginon/Dose/Dose Calibrator> Enter Screen Current to Beam Current Scale Factor as 1 as a starting point. |
||
32 | |
||
33 | # scope> Optionally lower the screen. Dose Calibrator will automatically do this if the user forgets. |
||
34 | |
||
35 | # Leginon/Dose/Dose Calibrator> Click Measure Dose Rate. The Dose Rate, Beam Current, and Screen Magnification will appear in the corresponding fields after they have been calculated. |
||
36 | |
||
37 | # Leginon/Dose/Dose Calibrator> Adjust the Screen Current to Beam Current Scale Factor so that the displayed beam current is equal to that of the Faraday cup measurement. NRAMM Tecnai's value is 0.88. |
||
38 | 1 | Amber Herold | |
39 | |||
40 | 3 | Amber Herold | |
41 | 1 | Amber Herold | h2. Calibrate the camera sensitivity |
42 | |||
43 | |||
44 | 3 | Amber Herold | |
45 | 4 | Amber Herold | # Scope> Move the stage to an empty area or a broken hole. |
46 | |
||
47 | # Scope> Expand the beam so that it covers the main screen and illuminate it uniformly. |
||
48 | |
||
49 | # Leginon/Dose/Settings> Configure camera with override preset activated. |
||
50 | |
||
51 | # Leginon/Dose/Dose Calibrator> |
||
52 | |
||
53 | 1 | Amber Herold | ** Beam Diameter = 0.16 m for the Tecnai's main viewing screen. |
54 | 4 | Amber Herold | |
55 | ** Screen Current to Beam Current Scale Factor = (as determined above or 1 to assume good calibration by the microscope manufacturer) |
||
56 | |
||
57 | # scope> Optionally lower the screen. Dose Calibrator will automatically do this if the user forgets. |
||
58 | |
||
59 | # Leginon/Dose/Dose Calibrator> Click Measure Dose Rate. The Dose Rate, Beam Current, and Screen Magnification will appear in the corresponding fields after they have been calculated. |
||
60 | |
||
61 | # Leginon/Dose/Dose Calibrator> Without changing the beam intensity, click on "Calibrate". An image is acquired by Leginon. |
||
62 | |
||
63 | 1 | Amber Herold | *Leginon will automatically lift the main viewing screen on the microscope. Some users report that the function does not work on their microscopes. Do it manually if the automatic function does not work. |
64 | 4 | Amber Herold | |
65 | # The calculated CCD sensitivity (counts / electron) will display once the calculation is completed. |
||
66 | 1 | Amber Herold | |
67 | 3 | Amber Herold | |
68 | 1 | Amber Herold | |
69 | |||
70 | h2. Enter pre-determined camera sensitivity value |
||
71 | |||
72 | 3 | Amber Herold | |
73 | 6 | Jim Pulokas | # Leginon/Dose/Calibrate/Camera Sensitivity> Enter Sensitivity values in counts per electron. |
74 | 4 | Amber Herold | |
75 | 6 | Jim Pulokas | # Leginon/Dose/Calibrate/Camera Sensitivity> Click on Save. |
76 | 2 | Amber Herold | |
77 | 5 | Jim Pulokas | !predetermined-dose.png! |
78 | |||
79 | 4 | Amber Herold | *Dose Calibration Need for the Example MSI:* |
80 | 2 | Amber Herold | |
81 | |_.Preset|_.magnification| |
||
82 | |*|any| |
||
83 | |||
84 | |||
85 | |||
86 | |||
87 | 1 | Amber Herold | ______ |
88 | |||
89 | [[Autofocus Calibration with Beam Tilt Node|< Autofocus Calibration with Beam Tilt Node]] |
||