Eagle camera installation and setup » History » Version 8
Anchi Cheng, 02/02/2016 02:32 PM
1 | 2 | Anchi Cheng | h1. TIA-controlled camera (FEI Eagle) |
---|---|---|---|
2 | 1 | Anchi Cheng | |
3 | h2. Extra Package and Installation |
||
4 | |||
5 | |_.program/package|_.notes| |
||
6 | |Tia.dll |(Should come with the microscope)| |
||
7 | |||
8 | {{include(comtypes installer)}} |
||
9 | |||
10 | 5 | Anchi Cheng | h3. Run checkcom.py |
11 | 1 | Anchi Cheng | |
12 | * From a command line window: |
||
13 | <pre>cd C:\python27\Lib\site-packages\pyscope |
||
14 | 6 | Anchi Cheng | C:\python27\python.exe checkcom.py</pre> |
15 | 1 | Anchi Cheng | |
16 | 5 | Anchi Cheng | * The output sould contain this output |
17 | 3 | Anchi Cheng | <pre> |
18 | 5 | Anchi Cheng | Found ESVision.Application done. |
19 | 1 | Anchi Cheng | </pre> |
20 | |||
21 | h2. instruments.cfg |
||
22 | |||
23 | Eagle: |
||
24 | <pre> |
||
25 | [camera] |
||
26 | 7 | Anchi Cheng | class: tia.TIA_Eagle |
27 | 1 | Anchi Cheng | zplane: 50 |
28 | height: 4096 |
||
29 | width: 4096 |
||
30 | </pre> |
||
31 | |||
32 | h2. Setup |
||
33 | |||
34 | * Set camera configuration to give the [[Leginon Image Orientation|standard Leginon orientation]]. |
||
35 | |||
36 | 4 | Anchi Cheng | h2. Programs to open before Leginon Client: TIA. |
37 | |||
38 | 1 | Anchi Cheng | h2. Testing with pyscope |
39 | |||
40 | In python command |
||
41 | <pre> |
||
42 | form pyscope import tia |
||
43 | 7 | Anchi Cheng | g = tia.TIA_Eagle() |
44 | 1 | Anchi Cheng | g.setExposureTime(200) |
45 | g.getImage() |
||
46 | </pre> |
||
47 | |||
48 | 8 | Anchi Cheng | {{include(pyscope_test_result array)}} |
49 | 1 | Anchi Cheng | |
50 | 4 | Anchi Cheng | h2. Trouble shooting |
51 | |||
52 | Some version of TIA may give an error of that at the end of it says |
||
53 | |||
54 | <pre> |
||
55 | _ctypes.COMError: (-2147352567, 'Exception occurred.', (u"Camera 'tia' does not exist.''.u'ESVision', None, 0L, 0)) |
||
56 | </pre> |
||
57 | This happens because TIA scripting (ESVision) knows the camera by a different name. |
||
58 | |||
59 | The correct name may be in TUI/CCDCamera panel. Or you can find out from these python commands when the camera in question is active (You can make it active by acquire an image in TIA). |
||
60 | <pre> |
||
61 | from pyscope import tia |
||
62 | t = tia.get_tiaccd() |
||
63 | c = t.ccd.Camera |
||
64 | </pre> |
||
65 | You should get the camera name as the output. |
||
66 | |||
67 | 7 | Anchi Cheng | With the name, edit pyscope/tia.py and change in the class TIA_Eagle, |
68 | 4 | Anchi Cheng | from |
69 | <pre> |
||
70 | 7 | Anchi Cheng | camera_name='BM-Eagle' |
71 | 4 | Anchi Cheng | </pre> |
72 | to |
||
73 | <pre> |
||