h1. Falcon 3 support
2
3
h2. Requirement from camera manufacture:
4
5
* Krios 2.9 or Talos/Arctica equivalent software version.
6
* Advanced TEM Scripting
7
8
h2. Required leginon version:
9
10
* 3.4 and above
11 2 Anchi Cheng
12
h2. Extra Package and Installation
13
14
|_.program/package|_.notes|
15 3 Anchi Cheng
|TEMAdvancedScripting.dll |(Should come with the microscope)|
16 2 Anchi Cheng
17
{{include(comtypes installer)}}
18
19
{{include(comtypes_installation)}}
20
21
You should have
22
23
<pre>
24
TEMAdvancedScripting
25
</pre>
26
27
h2. instruments.cfg
28
29
* Falcon3 class is for linear mode and Falcon3EC class is for electron counting mode. You want to have both.
30
<pre>
31
[camera]
32
class: feicam.Falcon3
33
zplane: 50
34
height: 4096
35
width: 4096
36
</pre>
37
<pre>
38
[camera]
39
class: feicam.Falcon3EC
40
zplane: 50
41
height: 4096
42
width: 4096
43
</pre>
44
45
h2. Setup
46
47
* Set dose rate and acquire References using TUI Reference manager.  Don't do it in Leginon.
48
* Set up in fei.cfg (See FEI scope installation instruction)
49
<pre>
50
[camera]
51
FRAME_SUBPATH = leginonframes
52
</pre>
53
In this case, you will need to create a directory called leginonframes on your Falcon image storage server (called OffloadData on your scope PC) with these parent directories
54
<pre>
55
OffloadData\TemScripting\BM-Falcon\leginonframes
56
</pre>
57
58
This leginonframes directory can then mounted on your permanent storage system and allow rawtransfer.py to rsync the files.
59
60
h2. Programs to open before Leginon Client: TIA.
61
62
h2. Testing with pyscope
63
64
In python command
65
<pre>
66
form pyscope import feicam
67
g = feicam.Falcon3()
68
g.setExposureTime(200)
69
g.getImage()
70
</pre>
71
72
{{include(pyscope_test_result array)}}
73
74
h2. Trouble shooting
75
76
If TEM Advanced Scripting is not setup correctly by the manufacture, a configuration error will appear similar to this:
77
<pre>
78
_ctypes.COMError: (-2147467259, 'Unspecified error.', (u"Could not load Storage Server configuration file: C:\\ProgramData\\FEI||AcquisitionStorageServer||StorageServer.config', u'',u'', 0, None))
79